Work Order 360 View

Detailed work order timeline, KPIs, and analysis

← WO Overview

WO-01274972 ↗ ServiceMax

GE Healthcare FMI • FCO • PC

📍 OHIOHEALTH RIVERSIDE METHODIST HOSPITAL — 3535 OLENTANGY RIVER RD, COLUMBUS OH, 43214

May 22, 2025 → Nov 17, 2025

✅ Completed
PC Per client agreement
Target: Jul 12, 2025
🕐 Clock stopped: Completed (Nov 17)
⏸ 18d excluded (parts wait + holding)
✗ SLA Missed (109d late)
Dispatch 0.3d
Holding 13.5d (excluded)
Coordinators 116.6d
Billing 11.3d
✗ SLA Missed
Dispatch
0.3 days
Holding (excluded from total)
13.5 days
Coordinators
116.6 days
Over SLA
Billing
11.3 days
Over SLA
Created: May 22, 2025 Due: Jul 12, 2025 Completed: Nov 17, 2025 (127 days late)
📅
128.0 days Total Age
⏱️
113.9d in Ready for Scheduling Longest Stage
🔄
13 transitions Status Changes
⚠️
122.1d over Total Delay
📦
0 orders, 0 quotes Parts & Quotes
📝
3 lines Work Details
🏠
1 visit On-Site Trips
📋
Coordinator Scheduling

AI Analysis

Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 357 days old
🚨
Primary Delay 113.9 days stuck in "Ready for Scheduling"
⏱️
Total Delay 122.1 business days over SLA thresholds

SLA vs Actual Time by Status

Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.3d / 0.5d SLA
Ready for Scheduling
115.9d / 1.0d SLA
Ready for Release
13.5d
Scheduled
0.7d / 2.0d SLA
Closed
1.1d / 1.0d SLA
Ready to Bill
11.3d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA

💬 Activities, Notes & Communications

🔒 Internal CSA Notes
09/29/2025 DMalloy - Emailed customer advising I will have 1-2 FSEs on site next week to begin the remediation on the 40+devices. Customer responding providing the details for AIS background instructions and agreed to the visit; email: brittani.penix@nationwidechildrens.org ... 07/23/2025 DMalloy - VM recieved from custoemr regarding the device with Serial Number: tabz72325. Customer advised that the device does have a broken heater door and will require the remediation. Called cst back at 614.566.3671. No answer left a VM asking if she could also make the other devices available as well for remediation on Next week Monday 7/28/2025 for FSE Paul to be onsite. Followed up with an email: brielle.koenig@ohiohealth.com ... 06/11/2025 DMalloy - Email received from customer advsing that they were going to allow EMSAR to complete the remediations on the 11 devices. Replied advsing that I can have FSE Paul on site on Monday, 6/30/2025; emails sent: jasmine.zirkle@nationwidechildrens.org ... brielle.koenig@ohiohealth.com ... ... 06/02/2025 DMalloy - Called recieved from Brielle Koenig 614.566.3671. She asked if we sent the part could she complete the recall herself. Advised I would look in to this and reached back out to her. Looped her in to the email, Brielle.Koenig@ohiohealth.com ... ... 06/02/2025 MHamersley - called 614-566-3727, call went to VM/LM; called 614-566-5000, transferred to Bio Med, call went to VM/LM. ... sent email: Tiffany.Miller@ohiohealth.com ; ... ... ** Email to Tiffany BOUNCED **
🔧 Work Performed
Template Name: FMI32097_Code 1_inspect. Template Description: FMI32097_Code 1_inspect. Problem Found: FMI32097_Heater door inspection. Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification. Verification Test: 1. Verified the canopy moves smoothly. 2. Verified transition mode message is not present. 3. Verified doors close smoothly. 4. Verified corner pins are aligned. 5. Verified no alarms. The device has passed all tests and is ready for customer use.
🔍 Technician Findings
nr
⚠️ Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄 Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞 Call Description
FMI 32097

Timeline

📋
Work Order Created May 22, 7:59 AM
Work Order: WO-01274972
Type: FCO
Priority: PC
📅
Status: Ready for Scheduling by Allison Belyea May 22, 10:25 AM
From: Entered
To: Ready for Scheduling
Duration in Previous: 2h 26m
2.0 business days
Status: Ready for Release by Mandy Newman May 23, 3:53 PM
From: Ready for Scheduling
To: Ready for Release
Duration in Previous: 1d 5h
🚨 14.0 business days
📅
Status: Ready for Scheduling by Mary Hamersley Jun 11, 12:31 PM
From: Ready for Release
To: Ready for Scheduling
Duration in Previous: 18d 20h
🚨 114.0 business days
📅
3 Dispatch Scheduled Nov 17, 12:41 PM
Dispatch Scheduled → Status: Scheduled → Technician Assigned
🗓️
Status: Scheduled by Deon Malloy
From: Ready for Scheduling
To: Scheduled
Duration in Previous: 159d 1h
👷
Technician Assigned by Deon Malloy
Technician: Paul Mccormish
📅
Dispatch Scheduled by Deon Malloy Nov 17, 12:54 PM
Scheduled For: 2025-11-17T13:00:00.000+0000
⏱️
2x Labor Added Nov 17, 5:45 PM
Type: Labor
Status: Open
Qty: 1, 0.75
✔️
Status: Completed by Paul Mccormish Nov 17, 5:54 PM
From: Scheduled, Completed
To: Completed, Closed
Duration in Previous: 5h 13m, 0m
⏱️
2 Labor Added Nov 18, 2:47 PM
Labor Added → Status: 07e-Partial Project
📋
Status: 07e-Partial Project by Timothy Amburgey
From: Closed, 07e-Partial Project
To: 07e-Partial Project, Ready to Bill
Duration in Previous: 20h 54m, 0m
4.0 business days
💵
2 Status: Billing Review Nov 21, 2:28 PM
Status: Billing Review → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From: Billing Review
To: Ready to Bill
Duration in Previous: 4m
3.0 business days
💵
2 Status: Billing Review Nov 25, 9:46 AM
Status: Billing Review → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From: Billing Review
To: Ready to Bill
Duration in Previous: 4m
⚠️ 5.0 business days
💵
2 Status: Billing Review Dec 1, 11:43 AM
Status: Billing Review → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From: Billing Review
To: Ready to Bill
Duration in Previous: 3m
📨
Status: Invoice Pending by Timothy Amburgey Dec 1, 1:40 PM
From: Ready to Bill, Invoice Pending
To: Invoice Pending, Invoiced
Duration in Previous: 1h 53m, 0m
⚠️ 6.0 business days
🧾
Invoice Created Dec 8, 4:35 PM
Invoice Number: 247861

Bottlenecks

Ready for Scheduling
Duration: 2.0 business days (threshold: 1.0 business days)
Exceeded by 1.0 business days
Ready for Scheduling
Duration: 113.9 business days (threshold: 1.0 business days)
Exceeded by 112.9 business days
Closed
Duration: 1.1 business days (threshold: 1.0 business days)
Exceeded by 1.1 business hours
Ready to Bill
Duration: 3.5 business days (threshold: 1.0 business days)
Exceeded by 2.5 business days
Ready to Bill
Duration: 2.6 business days (threshold: 1.0 business days)
Exceeded by 1.6 business days
Ready to Bill
Duration: 5.0 business days (threshold: 1.0 business days)
Exceeded by 4.0 business days

🔗 Related Artifacts

No related artifacts

ℹ️ Work Order Details

Status
Invoiced
Priority
PC
Type
FCO
Branch
N/A
Technician
Paul Mccormish
Scheduled
Nov 17, 8:00 AM
Created
May 22, 7:59 AM
Invoice #
247861

Work Details (3)

Name Type Status Qty Amount Date
WL-03010344 Labor Open 1 $0.00 Nov 17, 5:45 PM
WL-03010345 Labor Open 0.75 $100.00 Nov 17, 5:45 PM
WL-03011462 Labor Open 0.016666666666667 $38.00 Nov 18, 2:47 PM