01/06/2026 MHamersley - called 518-354-9108, spoke to Joe, he confirmed all 3 GE devices and ready to schedule the end of February.
12/04/2025 MHamersley - called 518-891-4141, xfer to Bio Med, call went to Joe's VM/LM; sent f/u email: javeni@uvm.edu
WO-01281092-Venue Go Ultrasound RVGB000481
WO-01282731-Venue Go Ultrasound VGD002694
WO-01281772-Venue Fit Ultrasound VFC000758
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R3.4 and communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”.
Verification Test:
“Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
5.
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101C: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📅
WO-01281772 - FCO - Venue Fit Ultrasound - SARANAC LAKE - THE ADIRONDACK MEDICAL CENTER
Feb 19, 2026 11:15am – 12:15pm
Work Order: WO-01281772
Account: GE Healthcare FMI
Customer Case Number: 1-554880082439, 1-554907754392
Location: THE ADIRONDACK MEDICAL CENTER
Address: 2233 NY-86, SARANAC LAKE, NY 12983
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFC000758
Contact Name: Joe Aveni
Contact Ph
JPG
Image - 2026-02-19T104342.027
Feb 19, 2026
Timeline
📋
Work Order Created Jun 19, 2:26 PM
Work Order:WO-01281772
Type:FCO
Priority:PC
✅
Status: Ready for Release by Allison BelyeaJun 19, 3:12 PM
From:Entered
To:Ready for Release
Duration in Previous:45m
🚨135.0 business days
📅
Status: Ready for Scheduling by Mary HamersleyJan 6, 11:07 AM