Created: Jun 19, 2025
Due: Feb 28, 2026
Completed: Feb 17, 2026 (11 days early)
📅
162.0 daysTotal Age
⏱️
99.0d in Ready for ReleaseLongest Stage
🔄
12 transitionsStatus Changes
⚠️
72.3d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 420 days old
🚨
Primary Delay99.0 days stuck in "Ready for Release"
⏱️
Total Delay72.3 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
99.0d
Scheduled
11.9d / 2.0d SLA
Ready for Scheduling
54.0d / 1.0d SLA
Tech On Site
0.0d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
4.0d / 1.0d SLA
Ready to Bill
9.5d / 1.0d SLA
Invoice Pending
0.3d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/09/2026 RFosbury - Jeffrey onsite Feb 16th and 17th
01/06/2026 MHamersley - will schedule when FSE and devices are available ... 11/19/2025 RFosbury - working with site to reschedule. Battery kits already onsite. ... 11/10/2025 RFosbury - confirmed with Ethan Bryant Rick C onsite Nov 20th and 21st ... 09/26/2025 Rfosbury - Sent email to Shawnell Nelson @ snelso29@jhmi.edu to confirm inventory. ... 08/19/2025 MHamersley - called 410-550-0265, call went to VM, wasn't accepting msgs; sent email: avelard1@jhmi.edu
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R4.4.3 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed
all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries /
Batteries installed by customer and dispose of old batteries
according to local regulation”.
Verification Test:
“Performed battery charging test - verified after 10 minutes, the scan
time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”"
🔍Technician Findings
GOOD WORKING CONDITIONS
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01282496
Account: GE Healthcare FMI
Customer Case Number: 1-554880334798, 1-554908083618
Location: JOHNS HOPKINS HOSPITAL MATERIALS MANAGEMENT
Address: 1780 E FAYETTE ST, BALTIMORE, MD 21287
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD002203
Contact Name: Eth
JPEG
VGB000203
Feb 16, 2026
JPEG
VGB000203_Diagnostic report
Feb 16, 2026
JPEG
VGB000203_New Battery
Feb 16, 2026
Timeline
📋
Work Order Created Jun 19, 2:27 PM
Work Order:WO-01282496
Type:FCO
Priority:PC
✅
Status: Ready for Release by Allison BelyeaJun 19, 3:15 PM