📍 CHRISTIANA HOSPITAL — 4755 OGLETOWN STANTON RD, NEWARK DE, 19718
Aug 25, 2025 → Nov 25, 2025
✅ Completed
PCPer client agreement
Target:Nov 30, 2025
🕐 Clock stopped: Completed (Nov 25)
⏸ 4d excluded (parts wait + holding)
✓ SLA Met (5d early)
Dispatch3.0d
Holding4.0d (excluded)
Coordinators63.3d
Billing4.0d
✓ SLA Met
Dispatch
3.0 days
Over SLA
Holding(excluded from total)
4.0 days
Coordinators
63.3 days
Over SLA
Billing
4.0 days
Over SLA
Created: Aug 25, 2025
Due: Nov 30, 2025
Completed: Nov 25, 2025 (5 days early)
📅
67.0 daysTotal Age
⏱️
42.0d in ScheduledLongest Stage
🔄
10 transitionsStatus Changes
⚠️
66.4d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
2 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 262 days old
🚨
Primary Delay42.0 days stuck in "Scheduled"
⏱️
Total Delay66.4 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
3.0d / 0.5d SLA
Ready for Scheduling
0.0d / 1.0d SLA
Scheduled
42.3d / 2.0d SLA
Ready for Release
4.0d
Assigned
21.0d / 0.5d SLA
Closed
1.4d / 1.0d SLA
07e-Partial Project
0.0d
Ready to Bill
4.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
11/25/2025 RFosbury dispatching to RICK C for today.
10/28/2025 Cever email sent to Carr, Lisa R to confirm new list of inventory and schedule FSE back onsite
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors
and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer
🔍Technician Findings
GOOD WORKING CONDITIONS
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01299345
Account: GE Healthcare FMI
Customer Case Number: 1-555761870294
Location: CHRISTIANA HOSPITAL
Address: 4755 OGLETOWN STANTON RD, NEWARK, DE 19718
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: HDGQ51096
Contact Name: Lisa Carr
Contact Phone: (302)7