Created: Sep 4, 2025
Due: Nov 30, 2025
Completed: Feb 12, 2026 (73 days late)
📅
109.0 daysTotal Age
⏱️
67.0d in Ready for SchedulingLongest Stage
🔄
8 transitionsStatus Changes
⚠️
103.2d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
5 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 343 days old
🚨
Primary Delay67.0 days stuck in "Ready for Scheduling"
⏱️
Total Delay103.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
17.0d
Ready for Scheduling
67.0d / 1.0d SLA
Awaiting Parts
27.0d / 3.0d SLA
Scheduled
1.0d / 2.0d SLA
Closed
1.2d / 1.0d SLA
Ready to Bill
14.0d / 1.0d SLA
Invoice Pending
0.4d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/12/2026 RFosbury dispatching to Paul for today.
09/26/2025 Mklose, per email from 09/09/2025 Brittani, Penix confirmed Devices, brittani.penix@nationwidechildrens.org ... 09/09/2025 MHamersley - called 614-355-1754, call went to Brittani's VM/LM; recd new POC info from GE: Brittani Penix, Medical Equipment Fleet Manager; sent email: brittani.penix@nationwidechildrens.org ... 08/28/2025 ABelyea - per Tabitha, ok to start calling on Nationwide now ... 07/17/2025 ABelyea - Per Tabitha Wright, placing FMI 32097 for this site On Hold until further notice ... 07/14/2025 MHamersley - called 614-234-7354, call went to Nolan's VM/LM; sent f/u email: nfuller2@mchs.com 07/09/2025 MHamersley - called 614-234-7354, call went to Nolan's VM/LM; sent f/u email: nfuller2@mchs.com 06/20/2025 MHamersley - called 614-234-7354, call went to Nolan's VM/LM; sent f/u email: nfuller2@mchs.com 06/10/2025 MHamersley - called 614-234-7354, call went to Nolan's VM/LM; sent f/u email: nfuller2@mchs.com 05/30/2025 MHamersley - called 614-234-7354, spoke to Nolan; sent email: nfuller2@mchs.com
🔧Work Performed
Template Name: FMI32097_Code 1_part_replace.
Template Description: FMI32097_Code 1_part_replace.
Problem Found: FMI32097_Heater door replacement.
Action Taken: FMI 32097 completed. Inspected the Heater doors. Heater doors found to be damaged. Replaced both heater doors as per Service Manual Addendum (5971733) and screws tightened as per instructions.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
nr
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.