📍 TRISTAR CENTENNIAL MEDICAL CENTER — 2300 PATTERSON ST, NASHVILLE TN, 37203
Sep 4, 2025 → Nov 21, 2025
✅ Completed
PCPer client agreement
Target:Nov 30, 2025
🕐 Clock stopped: Completed (Nov 19)
⏸ 25d excluded (parts wait + holding)
✓ SLA Met (11d early)
Dispatch0.0d
Holding18.0d (excluded)
Coordinators38.6d
Field Work2.9d
Billing6.0d
✓ SLA Met
Dispatch
0.0 days
Holding(excluded from total)
18.0 days
Coordinators
38.6 days
Over SLA
Field Work
2.9 days
Over SLA
Billing
6.0 days
Over SLA
Created: Sep 4, 2025
Due: Nov 30, 2025
Completed: Nov 21, 2025 (9 days early)
📅
57.0 daysTotal Age
⏱️
33.0d in Ready for SchedulingLongest Stage
🔄
7 transitionsStatus Changes
⚠️
42.7d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
12 linesWork Details
🏠
2 visitsOn-Site Trips
📋
CoordinatorScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
Visit 1Nov 19, 2025
Visit 2Nov 21, 2025
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 252 days old
🚨
Primary Delay33.0 days stuck in "Ready for Scheduling"
⏱️
Total Delay42.7 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
18.0d
Ready for Scheduling
33.0d / 1.0d SLA
Scheduled
5.6d / 2.0d SLA
Completed
2.9d / 1.0d SLA
Closed
1.2d / 1.0d SLA
Ready to Bill
6.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
11/12/2025 EMartin sent email to toby.breaux@gehealthcare.com advising FSE onsite Weds 11/19/2025 to complete devices
09/29/2025 Mklose called 615-342-1000, xfer to BioMed, Dylan, confirmed all devices except- Giraffe OmniBed Carestation TAB24240063SA .POC- Toby Breaux, toby.breaux@gehealthcare.com, dylan.moody@gehealthcare.com. sent vonfrimation email. Dylan last week, so Toby will be POC. ...09/12/2025 MHamersley - called 615-342-1000, xfer to Bio Med, call went to VM/LM
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors
and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
FMI needed
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📧
Email: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - TRISTAR CENTENNIAL MEDICAL CENTER - NASHVILLE, TN - WO-01301639
Completed Sep 29, 2025
To: dylan.moody@gehealthcare.com; toby.breaux@gehealthcare.com
CC: megan.klose@emsar.com; erin.martin@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: FMI32097 MIC UMDC English Customer Letter.pdf
Subject: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - TRISTAR CENTENNIAL MEDICAL CENTER - NASHVILLE, TN - WO-01301639
Body:
Hi Dylan, Toby
Thank you for taking my call today. As discussed, I am contacting you on behalf of GE Healthcare about a Field
📞
Live Call_Confirmed Devices During Call
Completed Sep 29, 2025
09/29/2025 Mklose called 615-342-1000, xfer to BioMed, Dylan, confirmed all devices except- Giraffe OmniBed Carestation TAB24240063SA
.POC- Toby Breaux, toby.breaux@gehealthcare.com, dylan.moody@gehealthcare.com. sent confirmation email. Dylan last week, so Toby will be POC.
📅
WO-01301639 - FCO - Giraffe OmniBed Carestation - NASHVILLE - TRISTAR CENTENNIAL MEDICAL CENTER
Nov 19, 2025 8:00am – 5:00pm
Work Order: WO-01301639
Account: GE Healthcare FMI
Customer Case Number: 1-555672029221
Location: TRISTAR CENTENNIAL MEDICAL CENTER
Address: 2300 PATTERSON ST, NASHVILLE, TN 37203
Manufacturer:
Model #: Giraffe OmniBed Carestation
Serial #: TAB24240065SA
Contact Name: Toby Breaux
Contact Phone: 615
PDF
TAB24240065SA_doors
Nov 19, 2025
PDF
TAB24240065SA_product label pic
Nov 19, 2025
PDF
TAB24240065SA_screws
Nov 19, 2025
Timeline
📋
Work Order Created Sep 4, 9:33 AM
Work Order:WO-01301639
Type:FCO
Priority:PC
✅
Status: Ready for Release by Allison BelyeaSep 4, 9:56 AM
From:Entered
To:Ready for Release
Duration in Previous:22m
🚨18.0 business days
📅
Status: Ready for Scheduling by Megan KloseSep 29, 11:51 AM