Created: Dec 12, 2025
Due: Mar 27, 2026
Completed: Mar 11, 2026 (16 days early)
📅
59.4 daysTotal Age
⏱️
49.9d in Ready for ReleaseLongest Stage
🔄
9 transitionsStatus Changes
⚠️
17.0d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 247 days old
🚨
Primary Delay49.9 days stuck in "Ready for Release"
⏱️
Total Delay17.0 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
49.9d
Ready for Scheduling
0.4d / 1.0d SLA
Scheduled
8.8d / 2.0d SLA
Tech On Site
1.2d / 1.0d SLA
Completed
0.1d / 1.0d SLA
Closed
4.0d / 1.0d SLA
Ready to Bill
6.6d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/26/2026 DMalloy - Emailed customer advsing that FSE Bennet will be onsite on Monday 3/9/2026 to complete the remediations on the 2 devices.
bodahn.kobuszewski@adventhealth.com
02/26/2026: ARoosa - Recvd. email from Bo stating they are ready to schedule and to provide him with a date and approx. ETA for tech and he will show tech where devices are located. Contact Bo upon arrival.
02/13/2026 MHamersley - sent f/u email: bodahn.kobuszewski@adventhealth.com
01/20/2026 MHamersley - called 913-676-7623, spoke to Bo, POC, FSE Cred: Vendormate; sent email: bodahn.kobuszewski@adventhealth.com
offering next week for FSE to come onsite.
WO-01325252-VGC000558
WO-01325253-VGD002918
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01325253
Account: GE Healthcare FMI
Customer Case Number: 1-554880155891, 1-554907997386
Location: ADVENTHEALTH SHAWNEE MISSION
Address: 9100 W 74TH ST, SHAWNEE MISSION, KS 66204
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD002918
Contact Name: Bodahn