📍 BAPTIST MEMORIAL HOSPITAL — 111 Dallas St, San Antonio TX, 78205
Dec 12, 2025 → Mar 29, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 31, 2026
✓ SLA Met (2 biz days early)
Active: 32 biz days
Paused: 39 biz days
🕐 Clock stopped: Completed (Mar 29, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸39d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 11, 2026
39 business days (not counted)
24d
▶ Clock Running
Status: Ready for Scheduling
Feb 11 → Mar 18, 2026
24 business days (counted)
Running total: 25 of ? biz days used
▶ Clock Running
Status: Scheduled
Mar 18 → Mar 24, 2026
4 business days (counted)
Running total: 29 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 24 → Mar 29, 2026
3 business days (counted)
Running total: 32 of ? biz days used
Created: Dec 12, 2025Completed: Mar 29, 2026
Show SLA Clock Detail (5 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 11
Ready for Release
⏸ Paused
39
—
Feb 11 → Mar 18
Ready for Scheduling
▶ Running
24
25
Mar 18 → Mar 24
Scheduled
▶ Running
4
29
Mar 24 → Mar 29
Tech On Site
▶ Running
3
32
Dispatch2.0d
Holding39.8d (excluded)
Coordinators29.1d
Field Work3.3d
Billing0.2d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
39.8 days
Coordinators
29.1 days
Over SLA
Field Work
3.3 days
Over SLA
Billing
0.2 days
Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 29, 2026 (2 days early)
📅
72.0 daysTotal Age
⏱️
39.8d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
30.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
22 linesWork Details
🏠
2 visitsOn-Site Trips
👤
Self-ScheduledScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
👤 Self-scheduled by Philmon Mehari — scheduling delays attributed to technician, not coordinators
Visit 1Mar 23, 2026
Visit 2Mar 24, 2026
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 246 days old
🚨
Primary Delay39.8 days stuck in "Ready for Release"
⏱️
Total Delay30.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
39.8d
Ready for Scheduling
24.1d / 1.0d SLA
Scheduled
5.0d / 2.0d SLA
Tech On Site
3.3d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
0.2d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
03/18/2026 mklose emailed Joseph adv Phil onsite March 24th.
02/11/2026 MHamersley - recd email reply from Joseph Golden, providing contact info for the other sites; He did confirm these 2 devices at his location: VGD000546 & VENGO000705
02/11/2026 MHamersley - returning missed call from Joseph Golden, 901-268-6991, sending him email, as he POC for several locations TN/MS/AR; email: joseph.golden@intelashealth.com
02/09/2026 MHamersley - called 901-226-5000, xfer to Bio Med, call went to VM/LM
Venue Go Ultrasound VENGO000705
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R2.8 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed
all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries /
Batteries installed by customer and dispose of old batteries
according to local regulation”.
Verification Test:“Performed battery charging test - verified after 10 minutes, the scan
time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”."
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R2.8 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed
all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries /
Batteries installed by customer and dispose of old batteries
according to local regulation”.
Verification Test:“Performed battery charging test - verified after 10 minutes, the scan
time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”."
⚠️Problem Description
FMI 78101A: Venue Go Thermal Event due to battery Purpose: Each affected system will have SW Version R2.9 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📞
Live Call_Left Message
Completed Feb 9, 2026
02/09/2026 MHamersley - called 901-226-5000, xfer to Bio Med, call went to VM/LM
📅
WO-01325288 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - MEMPHIS - BAPTIST MEMORIAL HOSPITAL
Mar 24, 2026 8:00am – 5:00pm
Work Order: WO-01325288
Account: GE Healthcare FMI
Customer Case Number: 1-554879664331, 1-554903852326
Location: BAPTIST MEMORIAL HOSPITAL
Address: 6019 WALNUT GROVE RD, MEMPHIS, TN 38120
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VENGO000705
Contact Name: Joseph Golden
Cont
JPG
VGB000705 _Product Label Pic
Mar 25, 2026
JPG
VGB000705 _Batteries
Mar 25, 2026
JPG
VGB000705 _Software Update
Mar 25, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325288
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨40.0 business days
📅
Status: Ready for Scheduling by Mary HamersleyFeb 11, 5:48 PM