📍 BAYLOR MEDICAL AT MCKINNEY — 5252 W UNIVERSITY DR, MCKINNEY TX, 75071
Dec 12, 2025 → Mar 12, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 31, 2026
✓ SLA Met (13 biz days early)
Active: 5 biz days
Paused: 55 biz days
🕐 Clock stopped: Completed (Mar 12, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸55d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Mar 6, 2026
55 business days (not counted)
▶ Clock Running
Status: Scheduled
Mar 6 → Mar 12, 2026
4 business days (counted)
Running total: 5 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 12 → Mar 12, 2026
0 business days (counted)
Running total: 5 of ? biz days used
Created: Dec 12, 2025Completed: Mar 12, 2026
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Mar 6
Ready for Release
⏸ Paused
55
—
Mar 6 → Mar 12
Scheduled
▶ Running
4
5
Mar 12 → Mar 12
Tech On Site
▶ Running
0
5
Dispatch2.0d
Holding55.8d (excluded)
Coordinators5.0d
Field Work0.0d
Billing6.1d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
55.8 days
Coordinators
5.0 days
Over SLA
Field Work
0.0 days
Billing
6.1 days
Over SLA
Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 12, 2026 (19 days early)
📅
61.0 daysTotal Age
⏱️
55.8d in Ready for ReleaseLongest Stage
🔄
7 transitionsStatus Changes
⚠️
12.3d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 247 days old
🚨
Primary Delay55.8 days stuck in "Ready for Release"
⏱️
Total Delay12.3 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
55.8d
Scheduled
5.0d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Closed
3.8d / 1.0d SLA
Ready to Bill
6.0d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
03/06/2026 DMalloy -Emailed customer advising that FSE Eric would be walking tin to this locaiton on next week friday 3/13/2026 for the remediation of the 5 devices at this location.
scsrecalls@bswhealth.org
03/05/2026 LHill-sent email to Amanda McCrory to see if she has a contact at this location. See logged email on WO-01325292.
03/04/2026 LHill-sent email to scsrecalls@bswhealth.org-Please see the email below. URGENT REQUEST. Need to get this scheduled ASAP.
See logged email on WO-01325292.
03/02/2026: ARoosa- Sent email to scsrecalls@bswhealth.org... 02/20/2026 MHamersley - called 512-259-6000, xfer to Temple Scott&White, xfer to Materials Mgmt, press 1, spoke to Missy, she provided: email: scsrecalls@bswhealth.org (this recall email address covers all of the Baylor Scott & White facilities in TX) ... 01/13/2026 MHamersley - called 469-764-1000, xfer to Bio Med, xfer to Ultrasound, need to call back tomorrow, 469-764-4434, ask for Connor or April
🔧Work Performed
Completed software only. Will return to replace battery once batteries are available
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📧
Email:BAYLOR MEDICAL AT MCKINNEY 5252 W UNIVERSITY DR MCKINNEY, TX 75071 WO-01325292
Completed Mar 5, 2026
From: laura.hill@emsar.com
To: amanda.mccrory@bswhealth.org
Cc: fmessak@emsar.com, rhea.berry@emsar.com
Good afternoon Amanda,
I'm trying to reach this location to schedule a tech to go onsite for a software update regarding the Venue Go Ultrasound devices.
Would you happen to have a contact person at the address below.
I’ve been having some difficulty getting in touch with anyone at that location, and any assistance or direction you could provide would be greatly appreciated.
Thank yo
📅
WO-01325292 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - MCKINNEY - BAYLOR MEDICAL AT MCKINNEY
Mar 13, 2026 9:00am – 6:00pm
Work Order: WO-01325292
Account: GE Healthcare FMI
Customer Case Number: 1-554880266335, 1-554908077119
Location: BAYLOR MEDICAL AT MCKINNEY
Address: 5252 W UNIVERSITY DR, MCKINNEY, TX 75071
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD000230
Contact Name:
Contact
PDF
VGD000230 COMPLETION REPORT
Mar 12, 2026
JPG
VGD000230 SERIAL NUMBER
Mar 12, 2026
JPG
VGD000230 NO WORKLIST
Mar 12, 2026
JPG
VGD000230 INFORMATION SCREEN
Mar 12, 2026
JPG
VGD000230 SOFTWARE
Mar 12, 2026
JPG
VGD000230 WIFI
Mar 12, 2026
JPG
VGD000230 BATTERIES
Mar 12, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325292
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:18 PM