📍 BAYLOR MEDICAL AT MCKINNEY — 5252 W UNIVERSITY DR, MCKINNEY TX, 75071
Dec 12, 2025 → Mar 11, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 31, 2026
✓ SLA Met (14 biz days early)
Active: 4 biz days
Paused: 55 biz days
🕐 Clock stopped: Completed (Mar 11, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸55d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Mar 6, 2026
55 business days (not counted)
▶ Clock Running
Status: Scheduled
Mar 6 → Mar 11, 2026
3 business days (counted)
Running total: 4 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 11 → Mar 11, 2026
0 business days (counted)
Running total: 4 of ? biz days used
Created: Dec 12, 2025Completed: Mar 11, 2026
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Mar 6
Ready for Release
⏸ Paused
55
—
Mar 6 → Mar 11
Scheduled
▶ Running
3
4
Mar 11 → Mar 11
Tech On Site
▶ Running
0
4
Dispatch2.0d
Holding55.8d (excluded)
Coordinators4.0d
Field Work0.4d
Billing6.6d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
55.8 days
Coordinators
4.0 days
Over SLA
Field Work
0.4 days
Billing
6.6 days
Over SLA
Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 11, 2026 (20 days early)
📅
60.0 daysTotal Age
⏱️
55.8d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
11.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 247 days old
🚨
Primary Delay55.8 days stuck in "Ready for Release"
⏱️
Total Delay11.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
55.8d
Scheduled
4.0d / 2.0d SLA
Tech On Site
0.4d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
3.8d / 1.0d SLA
Ready to Bill
6.6d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
03/06/2026 DMalloy -Emailed customer advising that FSE Eric would be walking tin to this locaiton on next week friday 3/13/2026 for the remediation of the 5 devices at this location.
scsrecalls@bswhealth.org
03/05/2026 LHill-sent email to Amanda McCrory to see if she has a contact at this location. See logged email on WO-01325292.
03/04/2026 LHill-sent email to scsrecalls@bswhealth.org-Please see the email below. URGENT REQUEST. Need to get this scheduled ASAP.
See logged email on WO-01325292.
03/02/2026: ARoosa- Sent email to scsrecalls@bswhealth.org... 02/20/2026 MHamersley - called 512-259-6000, xfer to Temple Scott&White, xfer to Materials Mgmt, press 1, spoke to Missy, she provided: email: scsrecalls@bswhealth.org (this recall email address covers all of the Baylor Scott & White facilities in TX) ... 01/13/2026 MHamersley - called 469-764-1000, xfer to Bio Med, xfer to Ultrasound, need to call back tomorrow, 469-764-4434, ask for Connor or April
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01325293 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - MCKINNEY - BAYLOR MEDICAL AT MCKINNEY
Mar 13, 2026 9:00am – 6:00pm
Work Order: WO-01325293
Account: GE Healthcare FMI
Customer Case Number: 1-554880397521, 1-554908085563
Location: BAYLOR MEDICAL AT MCKINNEY
Address: 5252 W UNIVERSITY DR, MCKINNEY, TX 75071
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD001617
Contact Name:
Contact
PDF
VGD001617 COMPLETION REPORT
Mar 11, 2026
JPG
VGD001617 SERIAL #
Mar 11, 2026
JPG
VGD001617 BATTERIES
Mar 11, 2026
JPG
VGD001617 WORKLIST
Mar 11, 2026
JPG
VGD001617 SOFTWARE
Mar 11, 2026
JPG
VGD001617 WIFI
Mar 11, 2026
JPG
VGD001617 INFORMATION REPORT
Mar 11, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325293
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:18 PM