📍 BAYLOR SCOTT & WHITE MEDICAL CENTER CENTENNIAL — 12505 LEBANON RD, FRISCO TX, 75035
Dec 12, 2025 → Mar 6, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 27, 2026
✓ SLA Met (15 biz days early)
Active: 7 biz days
Paused: 49 biz days
🕐 Clock stopped: Completed (Mar 6, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸49d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 26, 2026
49 business days (not counted)
▶ Clock Running
Status: Scheduled
Feb 26 → Mar 4, 2026
4 business days (counted)
Running total: 5 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 4 → Mar 6, 2026
2 business days (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Completed
Mar 6 → Mar 6, 2026
0 business days (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Closed
Mar 6 → Mar 6, 2026
0 business days (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Incomplete
Mar 6 → Mar 6, 2026
0 business days (counted)
Running total: 7 of ? biz days used
Created: Dec 12, 2025Completed: Mar 6, 2026
Show SLA Clock Detail (7 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 26
Ready for Release
⏸ Paused
49
—
Feb 26 → Mar 4
Scheduled
▶ Running
4
5
Mar 4 → Mar 6
Tech On Site
▶ Running
2
7
Mar 6 → Mar 6
Completed
▶ Running
0
7
Mar 6 → Mar 6
Closed
▶ Running
0
7
Mar 6 → Mar 6
Incomplete
▶ Running
0
7
Dispatch2.0d
Holding49.8d (excluded)
Coordinators4.9d
Field Work1.5d
Incomplete0.1d (excluded)
Billing10.0d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
49.8 days
Coordinators
4.9 days
Over SLA
Field Work
1.5 days
Over SLA
Incomplete(excluded from total)
0.1 days
Billing
10.0 days
Over SLA
Created: Dec 12, 2025
Due: Mar 27, 2026
Completed: Mar 6, 2026 (21 days early)
📅
57.0 daysTotal Age
⏱️
49.8d in Ready for ReleaseLongest Stage
🔄
9 transitionsStatus Changes
⚠️
16.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
9 linesWork Details
🏠
2 visitsOn-Site Trips
👤
Self-ScheduledScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
👤 Self-scheduled by Eric Mancerella — scheduling delays attributed to technician, not coordinators
Visit 1Mar 4, 2026
Visit 2Mar 5, 2026
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 247 days old
🚨
Primary Delay49.8 days stuck in "Ready for Release"
⏱️
Total Delay16.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
49.8d
Scheduled
4.9d / 2.0d SLA
Tech On Site
1.5d / 1.0d SLA
Closed
5.0d / 1.0d SLA
Incomplete
0.1d
Ready to Bill
10.0d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/26/2026 DMalloy - Dispatching for FSE Eric to Walk in to this location to locate and service the devices on next week Friday 3/6/2026.
02/20/2026 MHamersley - called 512-259-6000, xfer to Temple Scott&White, xfer to Materials Mgmt, press 1, spoke to Missy, she provided: email: scsrecalls@bswhealth.org (this recall email address covers all of the Baylor Scott & White facilities in TX)
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01325306 - FCO - DD: 03/27/2026 - Venue Go Ultrasound - FRISCO - BAYLOR SCOTT & WHITE MEDICAL CENTER CENTENNIAL
Mar 9, 2026 8:00am – 5:00pm
Work Order: WO-01325306
Account: GE Healthcare FMI
Customer Case Number: 1-555019962673, 1-554907750733
Location: BAYLOR SCOTT & WHITE MEDICAL CENTER CENTENNIAL
Address: 12505 LEBANON RD, FRISCO, TX 75035
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGC006107
Contact Na
PDF
VGC006107 COMPLETION REPORT
Mar 4, 2026
JPG
VGC006107 INFORMATION REPORT
Mar 4, 2026
JPG
VGC006107 SERIAL #
Mar 4, 2026
JPG
VGC006107 WIFI
Mar 4, 2026
JPG
VGC006107 NO WORKLIST
Mar 4, 2026
JPG
VGC006107 BATTERIES
Mar 4, 2026
JPG
VGC006107 SOFTWARE
Mar 5, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325306
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:18 PM