📍 BAYOU BEND HEALTH SYSTEMS — 1097 NORTHWEST BLVD, FRANKLIN LA, 70538
Dec 12, 2025 → Mar 14, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 31, 2026
✓ SLA Met (12 biz days early)
Active: 7 biz days
Paused: 54 biz days
🕐 Clock stopped: Completed (Mar 14, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸54d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Mar 5, 2026
54 business days (not counted)
6d
▶ Clock Running
Status: Scheduled
Mar 5 → Mar 14, 2026
6 business days (counted)
Running total: 7 of ? biz days used
Created: Dec 12, 2025Completed: Mar 14, 2026
Show SLA Clock Detail (3 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Mar 5
Ready for Release
⏸ Paused
54
—
Mar 5 → Mar 14
Scheduled
▶ Running
6
7
Dispatch2.0d
Holding54.9d (excluded)
Coordinators6.4d
Billing6.0d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
54.9 days
Coordinators
6.4 days
Over SLA
Billing
6.0 days
Over SLA
Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 14, 2026 (17 days early)
📅
62.0 daysTotal Age
⏱️
54.9d in Ready for ReleaseLongest Stage
🔄
7 transitionsStatus Changes
⚠️
11.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
6 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 247 days old
🚨
Primary Delay54.9 days stuck in "Ready for Release"
⏱️
Total Delay11.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
54.9d
Scheduled
6.4d / 2.0d SLA
Completed
0.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
6.0d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
03/05/206 mklose dispatch Davon-- unannounced
03/03/2026 YEspino Called 337-355-1250 to follow up with Jennifer over the previous email sent. Was unable to speak to her so i left a message.
02/27/2026: ARoosa - Called; 337-355-1250,
Jennifer Hudson Radiology Mngr. Wants me to send email with serial # to j.hudson@bayoubendhealth.org
02/25/2026: ARoosa- Called; 337-355-1240
Ray Debarbieris VM/LM sending email to
rdebarbieris@bayoubendhealth.org
02/19/2026: ARoosa - Called; 337-828-0760, asked to be xfer to biomed was sent to maintenance, asked if it was biomed he said yes it is part of biomed asked about device in question and was told I would need to contact biomed, I asked for a number or email and he had no idea of either.
01/29/2026 YEspino Called 337-828-0760 to verify the device in question. Asked to be transferred over to the biomed dept, Was unable to speak to Daniel Landry so i left a message.
🔧Work Performed
Completed software only. Will return to replace battery once batteries are available
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R2.8 or R2.9 and communicated with
customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
⚠️Problem Description
FMI 78101A: Venue Go Thermal Event due to battery Purpose: Each affected system will have SW Version R2.9 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📞
Live Call_ Left Message
Completed Mar 3, 2026
03/03/2026 YEspino Called 337-355-1250 to follow up with Jennifer over the previous email sent. Was unable to speak to her so i left a message.
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - BAYOU BEND HEALTH SYSTEMS - FRANKLIN, LA - WO-01325334
Completed Feb 27, 2026
To: j.hudson@bayoubendhealth.org
CC: angela.roosa@emsar.com; brandi.rodriguez@emsar.com; fmessak@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - BAYOU BEND HEALTH SYSTEMS - FRANKLIN, LA - WO-01325334
Body:
Hello Jennifer,
Thank you for returning my call today. As discussed, I am contacting you on behalf
📧
Live Call_Followed Up With Email
Completed Feb 27, 2026
02/27/2026: ARoosa - Called; 337-355-1250,
Jennifer Hudson Radiology Mngr. Wants me to send email with serial # to j.hudson@bayoubendhealth.org
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - BAYOU BEND HEALTH SYSTEMS - FRANKLIN, LA - WO-01325334
Completed Feb 25, 2026
To: rdebarbieris@bayoubendhealth.org
CC: angela.roosa@emsar.com; brandi.rodriguez@emsar.com; fmessak@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - BAYOU BEND HEALTH SYSTEMS - FRANKLIN, LA - WO-01325334
Body:
Hello Ray,
I am contacting you on behalf of GE Healthcare about a Field Safety Notice issued
📞
Live Call_Left Message
Completed Feb 25, 2026
02/25/2026: ARoosa- Called; 337-355-1240
Ray Debarbieris VM/LM sending email to
rdebarbieris@bayoubendhealth.org
📞
Live Call_Asked to Call Back
Completed Feb 19, 2026
02/19/2026: ARoosa - Called; 337-828-0760, asked to be xfer to biomed was sent to maintenance, asked if it was biomed he said yes it is part of biomed asked about device in question and was told I would need to contact biomed, I asked for a number or email and he had no idea of either.
📞
Live Call_ Left Message
Completed Jan 29, 2026
01/29/2026 YEspino Called 337-828-0760 to verify the device in question. Asked to be transferred over to the biomed dept, Was unable to speak to Daniel Landry so i left a message.
📅
WO-01325334 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - FRANKLIN - BAYOU BEND HEALTH SYSTEMS
Mar 12, 2026 12:00pm – 1:00pm
Work Order: WO-01325334
Account: GE Healthcare FMI
Customer Case Number: 1-554879749148, 1-554903819898
Location: BAYOU BEND HEALTH SYSTEMS
Address: 1097 NORTHWEST BLVD, FRANKLIN, LA 70538
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGB001870
Contact Name:
Contact Phone:
Con
JPG
VGB001870_No Battery in unit
Mar 14, 2026
JPG
VGB001870_Serial Label
Mar 14, 2026
JPG
VGB001870_No charge screen
Mar 14, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325334
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:11 PM