Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 29, 2026 (2 days early)
📅
72.0 daysTotal Age
⏱️
41.8d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
30.5d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
6 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 246 days old
🚨
Primary Delay41.8 days stuck in "Ready for Release"
⏱️
Total Delay30.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
41.8d
Ready for Scheduling
29.0d / 1.0d SLA
Scheduled
1.4d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
0.1d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/13/2026:ARoosa- Recvd. Email from Gelnna Scott confirming devices and is ready to schedule.
02/12/2026:ARoosa- Called; 731-593-6300, Xfer to Biomed> Glenna Scott, VM/LM, Recvd. call back sending email to glenna.scott@wth.org
02/12/2026:ARoosa- Called; 731-593-6300, Xfer to Biomed> Glenna Scott, VM/LM
01/22/2026 MHamersley/ARoosa - called 731-593-6300, xfer to Bio med, Glenna Scott, call went to VM/LM; called Amy Garner, HR, (731) 541-9914, call went to VM/LM
🔧Work Performed
Completed software upgrade only. Will return and replace batteries when they become available
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R3.4 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes.
Confirmed all probes pass. Completed all Peripheral Checks-system
passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries /
Batteries installed by customer and dispose of old batteries
according to local regulation”.
Verification Test:
“Performed battery charging test - verified after 10 minutes, the scan
time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully functional”."
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - CAMDEN GENERAL HOSPITAL - CAMDEN, TN - WO-01325353
Completed Feb 12, 2026
To: glenna.scott@wth.org
CC: angela.roosa@emsar.com; brandi.rodriguez@emsar.com; fmessak@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - CAMDEN GENERAL HOSPITAL - CAMDEN, TN - WO-01325353
Body:
Hello Glenna,
Thank you for taking my call today. As discussed, I am contacting you on behalf of GE Healthcare
📋
Inbound from +17315936300
Completed Feb 12, 2026
02/12/2026:ARoosa- Called; 731-593-6300, Xfer to Biomed> Glenna Scott, VM/LM, Recvd. call back sending email to glenna.scott@wth.org
📞
Live Call_Left Message
Completed Feb 12, 2026
02/12/2026:ARoosa- Called; 731-593-6300, Xfer to Biomed> Glenna Scott, VM/LM
📞
Live Call_ Left message
Completed Jan 22, 2026
01/22/2026 MHamersley/ARoosa - called 731-593-6300, xfer to Bio med, Glenna Scott, call went to VM/LM; called Amy Garner, HR, (731) 541-9914, call went to VM/LM
📞
Live Call_Left Message
Completed Jan 22, 2026
01/22/2026 MHamersley/ARoosa - called 731-593-6300, xfer to Bio med, Glenna Scott, call went to VM/LM; called Amy Garner, HR, (731) 541-9914, call went to VM/LM
📞
Live Call_Left Message
Completed Jan 22, 2026
01/22/2026 MHamersley/ARoosa - called 731-593-6300, xfer to Bio med, Glenna Scott, call went to VM/LM
📞
Live Call_No Answer Unable to Leave Message
Completed Jan 22, 2026
01/22/2026 MHamersley/ARoosa - called 731-593-6300, xfer to Radiology, transferred to wrong dept
📅
WO-01325353 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - CAMDEN - CAMDEN GENERAL HOSPITAL
Mar 27, 2026 10:00am – 11:00am
Work Order: WO-01325353
Account: GE Healthcare FMI
Customer Case Number: 1-554879876545, 1-554907665224
Location: CAMDEN GENERAL HOSPITAL
Address: 175 HOSPITAL DR, CAMDEN, TN 38320
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGC007481
Contact Name: Glenna Scott, Central Servic
JPG
VGC007481_Software Update
Mar 29, 2026
JPG
VGC007481_Product Label Pic
Mar 29, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325353
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨42.0 business days
📅
Status: Ready for Scheduling by Angela RoosaFeb 13, 8:29 AM