📍 ERLANGER MEDICAL CENTER — 910 BLACKFORD ST, CHATTANOOGA TN, 37403
Dec 12, 2025 → Mar 26, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 31, 2026
✓ SLA Met (2 biz days early)
Active: 32 biz days
Paused: 39 biz days
🕐 Clock stopped: Completed (Mar 27, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸35d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 5, 2026
35 business days (not counted)
▶ Clock Running
Status: Scheduled
Feb 5 → Feb 5, 2026
0 business days (counted)
Running total: 1 of ? biz days used
⏸ Clock Paused
Status: Awaiting Parts
Feb 5 → Feb 11, 2026
4 business days (not counted)
28d
▶ Clock Running
Status: Scheduled
Feb 11 → Mar 24, 2026
28 business days (counted)
Running total: 29 of ? biz days used
▶ Clock Running
Status: Entered
Mar 24 → Mar 24, 2026
0 business days (counted)
Running total: 29 of ? biz days used
▶ Clock Running
Status: Scheduled
Mar 24 → Mar 26, 2026
2 business days (counted)
Running total: 31 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 26 → Mar 26, 2026
0 business days (counted)
Running total: 31 of ? biz days used
▶ Clock Running
Status: Scheduled
Mar 26 → Mar 26, 2026
0 business days (counted)
Running total: 31 of ? biz days used
▶ Clock Running
Status: Tech On Site
Mar 26 → Mar 27, 2026
1 business day (counted)
Running total: 32 of ? biz days used
Created: Dec 12, 2025Completed: Mar 27, 2026
Show SLA Clock Detail (10 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 5
Ready for Release
⏸ Paused
35
—
Feb 5 → Feb 5
Scheduled
▶ Running
0
1
Feb 5 → Feb 11
Awaiting Parts
⏸ Paused
4
—
Feb 11 → Mar 24
Scheduled
▶ Running
28
29
Mar 24 → Mar 24
Entered
▶ Running
0
29
Mar 24 → Mar 26
Scheduled
▶ Running
2
31
Mar 26 → Mar 26
Tech On Site
▶ Running
0
31
Mar 26 → Mar 26
Scheduled
▶ Running
0
31
Mar 26 → Mar 27
Tech On Site
▶ Running
1
32
Dispatch2.0d
Material Management4.6d
Holding35.8d (excluded)
Coordinators31.6d
Field Work1.2d
Billing0.6d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Material Management
4.6 days
Over SLA
Holding(excluded from total)
35.8 days
Coordinators
31.6 days
Over SLA
Field Work
1.2 days
Over SLA
Billing
0.6 days
Created: Dec 12, 2025
Due: Mar 31, 2026
Completed: Mar 27, 2026 (4 days early)
📅
71.2 daysTotal Age
⏱️
35.8d in Ready for ReleaseLongest Stage
🔄
12 transitionsStatus Changes
⚠️
32.9d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
7 linesWork Details
🏠
2 visitsOn-Site Trips
👤
Self-ScheduledScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
👤 Self-scheduled by Ryan Lewis — scheduling delays attributed to technician, not coordinators
Visit 1Mar 26, 2026
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 246 days old
🚨
Primary Delay35.8 days stuck in "Ready for Release"
⏱️
Total Delay32.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
35.8d
Scheduled
31.6d / 2.0d SLA
Awaiting Parts
4.6d / 3.0d SLA
Tech On Site
1.2d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
3.0d / 1.0d SLA
Ready to Bill
0.6d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/04/2026 Brod - called 2103316429 called to speak to Randy, he stated i will have to work with Richard Branch to coordinate a time , he is going to respond back to my email with Richard to continue to schedule.
01/12/2026 MHamersley - called 423-778-7000, xfer to Bio Med, ext 2063, call went to VM/LM
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
GEHC FMI 78101 WOM Process
Published: 06/12/2025 Page 7 of 14 v 1.0
Effective: 06/12/2025
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
FMI
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01325371 - FCO - DD: 3/31/2026 - - Venue Go Ultrasound - CHATTANOOGA - ERLANGER MEDICAL CENTER
Mar 25, 2026 8:00pm – 8:00pm
Work Order: WO-01325371
Account: GE Healthcare FMI
Customer Case Number: 1-554880374394, 1-554908085526
Location: ERLANGER MEDICAL CENTER
Address: 910 BLACKFORD ST, CHATTANOOGA, TN 37403
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD003427
Contact Name: Richard Brannon
Contac
JPG
VGD003427_Batteries
Mar 26, 2026
JPG
VGD003427_SN
Mar 26, 2026
JPG
VGD003427_Software
Mar 26, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325371
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨36.0 business days
👷
▶
5
Technician Assigned Feb 5, 2:07 PM
Technician Assigned → Dispatch Scheduled → Status: Scheduled → Parts Requested → Status: Awaiting Parts