Created: Dec 12, 2025
Due: Feb 28, 2026
Completed: Feb 24, 2026 (4 days early)
📅
48.2 daysTotal Age
⏱️
26.4d in ScheduledLongest Stage
🔄
8 transitionsStatus Changes
⚠️
32.2d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
6 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 244 days old
🚨
Primary Delay26.4 days stuck in "Scheduled"
⏱️
Total Delay32.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
20.8d
Scheduled
26.4d / 2.0d SLA
Tech On Site
0.6d / 1.0d SLA
Completed
0.6d / 1.0d SLA
Closed
1.0d / 1.0d SLA
Ready to Bill
7.3d / 1.0d SLA
Invoice Pending
0.3d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/23/2026 SRobinson - Created follow up WO as requested. (WO-01353392)
02/23/2026 BBallard - Software updated. Needs follow up once batteries are ordered.
01/14/2026 Brod - called 8653424628 to speak to Yolanda, went to vm left message to have FSE onsite for 01/16. following up in email to yburkhart@endoconsultants.com
01/14/2026 Brod - recieved email with POC contact for The system is located at:
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📞
Live Call_ Left Message
Completed Jan 14, 2026
01/14/2026 Brod - called 8653424628 to speak to Yolanda, went to vm left message to have FSE onsite for 01/16. following up in email to yburkhart@endoconsultants.com
Feb 9, 2026 7:36am
Hi @Bruce Ballard Was this completed, or does this need to be rescheduled?
@Brandi Rodriguez
📅
WO-01325468 - FCO - Venue Fit Ultrasound - KNOXVILLE - ENDOCRINOLOGY CONSULTANTS OF
Feb 22, 2026 7:00pm – 7:00pm
Work Order: WO-01325468
Account: GE Healthcare FMI
Customer Case Number: 1-554880582377, 1-554907908104
Location: ENDOCRINOLOGY CONSULTANTS OF
Address: 1450 DOWELL SPRINGS BLVD STE 300, KNOXVILLE, TN 37909
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFD001100
Contact Name: Co
JFIF
VFD001100 SW
Feb 23, 2026
JFIF
VFD001100 SN
Feb 23, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325468
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM