Created: Dec 12, 2025
Due: Jan 31, 2026
Completed: Jan 13, 2026 (18 days early)
📅
23.0 daysTotal Age
⏱️
12.9d in Ready for ReleaseLongest Stage
🔄
9 transitionsStatus Changes
⚠️
21.7d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 153 days old
🚨
Primary Delay12.9 days stuck in "Ready for Release"
⏱️
Total Delay21.7 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
12.9d
Scheduled
0.1d / 2.0d SLA
Awaiting Parts
7.0d / 3.0d SLA
Tech Off Site
1.6d / 2.0d SLA
Completed
2.0d / 1.0d SLA
Closed
4.8d / 1.0d SLA
Ready to Bill
12.5d / 1.0d SLA
Invoice Pending
1.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/31/2025 EMartin sent email to william.horvat@gehealthcare.com advising FSE onsite tues 01/06/26 to start devices
12/30/2025 MHamersley - called 954-540-9821, spoke to Shawn, they are Bio Med for Westside Hospital only, include William Horvat in emails; william.horvat@gehealthcare.com ... 12/22/2025 Cever called 954-540-9821, LVM trying to confirm both locations and inventory along with schedule FSE onsite asap. Sending email to Shawn.Doolittle@gehealthcare.com ... 12/16/2025 Cever sent email to Doolittle, Shawn with inventory list for HCA Northwest and HCA Westside
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01325609
Account: GE Healthcare FMI
Customer Case Number: 1-554879960075, 1-554907742962
Location: HCA FLORIDA WESTSIDE HOSPITAL
Address: 8201 W BROWARD BLVD, PLANTATION, FL 33324
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGC006967
Contact Name:
Contact Phone
JPG
VGC006967_Product Label Pic
Jan 10, 2026
JPG
VGC006967_Product Pic
Jan 10, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325609
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:10 PM