12/31/2025 MHamersley - called 832-534-5000, busy signal x2; called 832-534-7800, press 8, this is for Primary care, need to call Imaging, operator couldn't provide assistance; called 832-539-3073, press 4, not right extension, this is for scheduling and operator wasn't helpful
WO-01326095
WO-01326096
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
FMI implemented.
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📧
Email
Completed Feb 4, 2026
02/04/2026; ARoosa Sent F/U email to> maria.terrazascruz@commonspirit.org
📞
Live Call_Asked to Call Back
Completed Dec 31, 2025
12/31/2025 MHamersley - called 832-534-5000, busy signal x2; called 832-534-7800, press 8, this is for Primary care, need to call Imaging, operator couldn't provide assistance
📞
Live Call_No Answer Unable to Leave Message
Completed Dec 31, 2025
12/31/2025 MHamersley - called 832-534-5000, busy signal
📅
WO-01326096 - FCO - HOUSTON - ST LUKES HOSPITAL
Feb 16, 2026 7:00pm – 7:00pm
Work Order: WO-01326096
Account: GE Healthcare FMI
Customer Case Number: 1-554879971064, 1-554907803809
Location: ST LUKES HOSPITAL
Address: 20171 CHASEWOOD PARK DR, HOUSTON, TX 77070
Manufacturer:
Model #:
Serial #: RVGC006332
Contact Name:
Contact Phone:
Contact Email:
Due Date: 2/28/2026
Sch
PDF
VGC006332_Self Test Post FMI
Feb 9, 2026
PDF
VGC006332_About System Version 1
Feb 9, 2026
PDF
VGC006332_Product Label Pic
Feb 9, 2026
PDF
VGC006332_Admin System Admin 1
Feb 9, 2026
PDF
VGC006332_Initial battery status
Feb 17, 2026
PDF
VGC006332_Self Test HW Change New Batteries
Feb 17, 2026
PDF
VGC006332_battery status after 30+ minutes
Feb 17, 2026
PDF
VGC006332_Batteries
Feb 17, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326096
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:16 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨38.0 business days
📅
Status: Ready for Scheduling by Angela RoosaFeb 4, 3:01 PM
From:Ready for Release
To:Ready for Scheduling
Duration in Previous:51d 2h
📅
▶
5
Dispatch Scheduled Feb 5, 11:10 AM
Dispatch Scheduled → Status: Scheduled → Technician Assigned → Parts Requested → Status: Awaiting Parts
🗓️
Status: Scheduled by Brandi Rodriguez
From:Ready for Scheduling
To:Scheduled
Duration in Previous:20h 9m
👷
Technician Assigned by Brandi Rodriguez
Technician:Marvin Rivas
📦
Parts Requested
Order:00089780
Status:Closed
📦
Status: Awaiting Parts by Brandi Rodriguez
From:Scheduled
To:Awaiting Parts
Duration in Previous:1m
⏳3.0 business days
🔧
Status: Tech On Site by Marvin RivasFeb 9, 10:31 AM
From:Awaiting Parts
To:Tech On Site
Duration in Previous:3d 23h
⏱️
▶
2
2x Labor Added Feb 9, 8:31 PM
2x Labor Added → Travel Added
🚗
Travel Added
Type:Travel
Status:Open
Qty:15
⚠️
Status: Incomplete by Marvin RivasFeb 9, 8:51 PM
From:Tech On Site
To:Incomplete
Duration in Previous:10h 20m
⚠️6.4 business days
🔧
Status: Tech On Site by Marvin RivasFeb 17, 10:38 AM
From:Incomplete
To:Tech On Site
Duration in Previous:7d 13h
🗓️
▶
2
Status: Scheduled Feb 17, 2:13 PM
Status: Scheduled → Dispatch Scheduled
📅
Dispatch Scheduled by Marvin Rivas
Scheduled For:2026-02-17T14:00:00.000+0000
⏱️
3x Labor Added Feb 17, 11:09 PM
Type:Labor, Travel
Status:Open
Qty:1.5, 15
✔️
▶
2
Status: Completed Feb 17, 11:40 PM
Status: Completed → Status: Closed
🏁
Status: Closed by Marvin Rivas
From:Completed
To:Closed
Duration in Previous:4m
⏳3.0 business days
⏱️
▶
2
Labor Added Feb 20, 2:57 PM
Labor Added → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From:Closed
To:Ready to Bill
Duration in Previous:2d 15h
⚠️5.0 business days
💵
▶
2
Status: Billing Review Feb 26, 2:31 PM
Status: Billing Review → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From:Billing Review
To:Ready to Bill
Duration in Previous:2m
⚠️6.0 business days
📨
Status: Invoice Pending by Timothy AmburgeyMar 5, 8:54 AM
From:Ready to Bill
To:Invoice Pending
Duration in Previous:6d 18h
🧾
▶
2
Invoice Created Mar 5, 11:18 AM
Invoice Created → Status: Invoiced
💰
Status: Invoiced by Timothy Amburgey
From:Invoice Pending
To:Invoiced
Duration in Previous:2h 23m
Bottlenecks
Entered
Duration: 2.0 business days (threshold: 4.0 business hours)
Exceeded by 1.5 business days
Ready for Scheduling
Duration: 1.5 business days (threshold: 1.0 business days)
Exceeded by 4.0 business hours
Tech On Site
Duration: 1.2 business days (threshold: 1.0 business days)
Exceeded by 1.9 business hours
Closed
Duration: 3.0 business days (threshold: 1.0 business days)
Exceeded by 2.0 business days
Ready to Bill
Duration: 4.5 business days (threshold: 1.0 business days)
Exceeded by 3.5 business days
Ready to Bill
Duration: 5.6 business days (threshold: 1.0 business days)