📍 PALEY ORTHOPEDIC & SPINE INSTITUTE — 901 45TH ST, MANGONIA PARK FL, 33407
Dec 12, 2025 → Jan 13, 2026
✅ Completed
PCPer client agreement
Target:Jan 31, 2026
🕐 Clock stopped: Tech Off Site (Jan 9)
⏸ 24d excluded (parts wait + holding)
✓ SLA Met (22d early)
Dispatch2.0d
Material Management8.0d
Holding12.9d (excluded)
Coordinators0.5d
Field Work2.2d
Billing14.1d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Material Management
8.0 days
Over SLA
Holding(excluded from total)
12.9 days
Coordinators
0.5 days
Field Work
2.2 days
Over SLA
Billing
14.1 days
Over SLA
Created: Dec 12, 2025
Due: Jan 31, 2026
Completed: Jan 13, 2026 (18 days early)
📅
23.0 daysTotal Age
⏱️
12.9d in Ready for ReleaseLongest Stage
🔄
12 transitionsStatus Changes
⚠️
23.2d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 153 days old
🚨
Primary Delay12.9 days stuck in "Ready for Release"
⏱️
Total Delay23.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
12.9d
Scheduled
0.5d / 2.0d SLA
Awaiting Parts
8.0d / 3.0d SLA
Tech Off Site
0.2d / 2.0d SLA
Completed
2.0d / 1.0d SLA
Closed
5.8d / 1.0d SLA
Ready to Bill
13.0d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
Invoice Pending
1.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/31/2025 EMartin called (561) 844-5255 / transferred to front desk placed on hold spk with S.Mancino confirmed onsite for Thrs 01/08/26 smancino@paleyinstitute.org ext 293
12/29/2025 MHamersley - called 561-882-2725, no answer/unable to leave msg; sent f/u email: Ronald.Doropan@tenethealth.com
(we have another site scheduled for Jan 5 with same POC) ... 12/23/2025 Cever sent email to ronald.doropan@tenethealth.com trying to confirm inventory
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📅
WO-01326098 - FCO - MANGONIA PARK - PALEY ORTHOPEDIC & SPINE INSTITUTE
Jan 9, 2026 8:00am – 5:00pm
Work Order: WO-01326098
Account: GE Healthcare FMI
Customer Case Number: 1-555020926242, 1-554907622453
Location: PALEY ORTHOPEDIC & SPINE INSTITUTE
Address: 901 45TH ST, MANGONIA PARK, FL 33407
Manufacturer:
Model #:
Serial #: RVGB000402
Contact Name:
Contact Phone:
Contact Email:
Due Date: 1/
JPG
VGB000402_Product Label Pic
Jan 10, 2026
JPG
VGB000402_Software Update
Jan 10, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326098
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:10 PM