📍 UNIVERSITY MEDICAL CENTER — 602 INDIANA AVE, LUBBOCK TX, 79415
Dec 12, 2025 → Feb 21, 2026
✅ Completed
PCPer client agreement
Target:Feb 28, 2026
🕐 Clock stopped: Completed (Feb 21)
⏸ 59d excluded (parts wait + holding)
✓ SLA Met (7d early)
Dispatch2.0d
Holding43.8d (excluded)
Coordinators4.4d
Field Work2.5d
Billing9.2d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
43.8 days
Coordinators
4.4 days
Over SLA
Field Work
2.5 days
Over SLA
Billing
9.2 days
Over SLA
Created: Dec 12, 2025
Due: Feb 28, 2026
Completed: Feb 21, 2026 (7 days early)
📅
51.0 daysTotal Age
⏱️
43.8d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
13.4d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
7 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 153 days old
🚨
Primary Delay43.8 days stuck in "Ready for Release"
⏱️
Total Delay13.4 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
43.8d
Scheduled
4.4d / 2.0d SLA
Tech On Site
2.5d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
9.0d / 1.0d SLA
Invoice Pending
0.2d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/23/2026 SRobinson- Created follow up WO as requested. (WO-01353282)
02/12/2026 RFosbury dispatching to Eric M for TR. Will dispatch others on demand next week.
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R3.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass.Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📅
WO-01326195 - FCO - LUBBOCK - UNIVERSITY MEDICAL CENTER
Feb 17, 2026 7:00pm – 7:00pm
Work Order: WO-01326195
Account: GE Healthcare FMI
Customer Case Number: 1-555020947311, 1-554907624115
Location: UNIVERSITY MEDICAL CENTER
Address: 602 INDIANA AVE, LUBBOCK, TX 79415
Manufacturer:
Model #:
Serial #: VGB002566
Contact Name:
Contact Phone:
Contact Email:
Due Date: 2/28/2026
Sche
JPG
VGB002566 INFORMATION REPORT
Feb 18, 2026
JPG
VGB002566 UPGRADED FROM 2 TO 3
Feb 18, 2026
PDF
VGB002566 COMPLETION REPORT
Feb 18, 2026
JPG
VGB002566 SERIAL #
Feb 18, 2026
JPG
VGB002566 NO WIFI
Feb 18, 2026
JPG
VGB002566 BATTERIES
Feb 18, 2026
JPG
VGB002566 NO WORKLIST
Feb 18, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326195
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:19 PM