📍 UNIVERSITY MEDICAL CENTER — 602 INDIANA AVE, LUBBOCK TX, 79415
Dec 12, 2025 → Feb 19, 2026
✅ Completed
PCPer client agreement
Target:Feb 28, 2026
🕐 Clock stopped: Completed (Feb 19)
⏸ 66d excluded (parts wait + holding)
✓ SLA Met (9d early)
Dispatch2.0d
Holding48.8d (excluded)
Coordinators0.1d
Field Work0.1d
Billing9.0d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
48.8 days
Coordinators
0.1 days
Field Work
0.1 days
Billing
9.0 days
Over SLA
Created: Dec 12, 2025
Due: Feb 28, 2026
Completed: Feb 19, 2026 (9 days early)
📅
50.0 daysTotal Age
⏱️
48.8d in Ready for ReleaseLongest Stage
🔄
9 transitionsStatus Changes
⚠️
10.3d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 153 days old
🚨
Primary Delay48.8 days stuck in "Ready for Release"
⏱️
Total Delay10.3 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
48.8d
Scheduled
0.1d / 2.0d SLA
Tech On Site
0.1d / 1.0d SLA
Closed
3.1d / 1.0d SLA
Ready to Bill
8.7d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
Invoice Pending
0.3d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/20/2026 SRobinson- Created follow up WO as requested. (WO-01353010)
02/13/2026 MHamersley - called 806-775-8200, xfer to Bio Med, call went to VM/LM
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📅
WO-01326200 - FCO - LUBBOCK - UNIVERSITY MEDICAL CENTER
Feb 19, 2026 8:00am – 5:00pm
Work Order: WO-01326200
Account: GE Healthcare FMI
Customer Case Number: 1-554880282777, 1-554908118702
Location: UNIVERSITY MEDICAL CENTER
Address: 602 INDIANA AVE, LUBBOCK, TX 79415
Manufacturer:
Model #:
Serial #: VGD000236
Contact Name:
Contact Phone:
Contact Email:
Due Date: 2/28/2026
Sche
JPG
VGD000236 SOFTWARE
Feb 19, 2026
JPG
VGD000236 SERIAL #
Feb 19, 2026
JPG
VGD000236 BATTERIES
Feb 19, 2026
JPG
VGD000236 NO WORKLIST
Feb 19, 2026
PDF
VGD000236 COMPLETION REPORT
Feb 19, 2026
JPG
VGD000236 NO WORKLIST
Feb 19, 2026
JPG
VGD000236 SERIAL #
Feb 19, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326200
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:16 PM