Created: Dec 12, 2025
Due: Jan 31, 2026
Completed: Jan 9, 2026 (22 days early)
📅
21.0 daysTotal Age
⏱️
10.8d in Ready for ReleaseLongest Stage
🔄
10 transitionsStatus Changes
⚠️
26.6d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
6 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 153 days old
🚨
Primary Delay10.8 days stuck in "Ready for Release"
⏱️
Total Delay26.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
10.8d
Scheduled
8.7d / 2.0d SLA
Awaiting Parts
0.4d / 3.0d SLA
Tech On Site
1.9d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
8.8d / 1.0d SLA
Ready to Bill
10.7d / 1.0d SLA
Invoice Pending
1.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/29/2025 EMartin rcvd email from POC asking when remaining devices will be completed replied advising FSE onsite 01/05/26 after holidays to complete remaining devices
12/15/2025 Cever sent email to Victor offering FSE onsite beginning Dec 22nd victor.cano1@gehealthcare.com 12/10/2025 MHamersley - called 713-790-1234, xfer to Bio Med, spoke to Victor Cano, POC, 262-449-3720; Location address is correct; FSE Cred: None; sent email: victor.cano1@gehealthcare.com
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
Implemented FMI.
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📅
WO-01326321 - FCO - HOUSTON - WOMANS HOSPITAL OF TEXAS
Jan 5, 2026 8:00am – 5:00pm
Work Order: WO-01326321
Account: GE Healthcare FMI
Customer Case Number: 1-554880303962, 1-554908095047
Location: WOMANS HOSPITAL OF TEXAS
Address: 7600 FANNIN ST, HOUSTON, TX 77054-1906
Manufacturer:
Model #:
Serial #: VGD001381
Contact Name:
Contact Phone:
Contact Email:
Due Date: 1/31/2026
S
PDF
VGD001381_Batteries
Jan 9, 2026
PDF
VGD001381_Product Label Pic
Jan 9, 2026
PDF
VGD001381 SERVICE REPORT POST FMI 78101_QR code email to self
Jan 9, 2026
PDF
VGD001381_Self Test Post FMI
Jan 9, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326321
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:18 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨11.0 business days
👷
▶
5
Technician Assigned Dec 29, 9:41 AM
Technician Assigned → Status: Scheduled → Dispatch Scheduled → Parts Requested → Status: Awaiting Parts