Work Order 360 View
Detailed work order timeline, KPIs, and analysis
P5
5 business days from creation
Target:
Jan 19, 2026
🕐 Clock stopped: Completed (Jan 12)
✓ SLA Met (6d early)
Coordinators
0.7 days
Over SLA
Created: Jan 12, 2026
Due: Jan 19, 2026
Completed: Jan 12, 2026 (7 days early)
⏱️
1.3d in Closed
Longest Stage
🔄
6 transitions
Status Changes
📦
0 orders, 0 quotes
Parts & Quotes
AI Analysis
Powered by OpenAI
Generating AI analysis...
🚨
Primary Delay
1.3 days stuck in "Closed"
⏱️
Total Delay
0.4 business days over SLA thresholds
SLA vs Actual Time by Status
Ready for Scheduling
0.1d
/ 1.0d SLA
Ready to Bill
0.0d
/ 1.0d SLA
Invoice Pending
0.0d
/ 2.0d SLA
💬 Activities, Notes & Communications
🔧
Work Performed
reset cam guide - tightened fowler link - reset pins in footboard connector - cycle tested unit - returned to service
🔍
Technician Findings
found head of bed would not raise all the way - found footboard inop
⚠️
Problem Description
repair request
head of bed function inop
Sebzda, Matthew
📄
Description
Bed, Electric
Timeline
📋
📅
From:
Entered
To:
Ready for Scheduling
Duration in Previous:
0m
👷
👷
From:
Ready for Scheduling
To:
Assigned
Duration in Previous:
53m
⏱️
✔️
From:
Assigned, Completed
To:
Completed, Closed
Duration in Previous:
4h 33m, 0m
📄
From:
Closed
To:
Ready to Bill
Duration in Previous:
21h 20m
📨
💰
From:
Invoice Pending
To:
Invoiced
Duration in Previous:
2m
🧾
Invoice Number:
SM-251086
Bottlenecks
Assigned
Duration: 4.6 business hours (threshold: 4.0 business hours)
Exceeded by 0.6 business hours
Closed
Duration: 1.3 business days (threshold: 1.0 business days)
Exceeded by 2.5 business hours
ℹ️ Work Order Details
Technician
Matthew Sebzda
Work Details (1)
| Name |
Type |
Status |
Qty |
Amount |
Date |
| WL-03062087 |
Labor |
Open |
1.5 |
$0.00 |
Jan 12, 4:29 PM |