📍 CARL R DARNALL ARMY MEDICAL CENTER — 36065 SANTA FE AVE, FORT HOOD TX, 76544
Jan 28, 2026 → Apr 13, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Apr 27, 2026
✓ SLA Met (9 biz days early)
Active: 12 biz days
Paused: 41 biz days
🕐 Clock stopped: Completed (Apr 14, 2026)
▶ Clock Running
Status: Created
Jan 28 → Jan 28, 2026
0 business days (counted)
Running total: 0 of ? biz days used
⏸41d
⏸ Clock Paused
Status: Ready for Release
Jan 28 → Mar 27, 2026
41 business days (not counted)
8d
▶ Clock Running
Status: Ready for Scheduling
Mar 27 → Apr 8, 2026
8 business days (counted)
Running total: 8 of ? biz days used
▶ Clock Running
Status: Assigned
Apr 8 → Apr 8, 2026
0 business days (counted)
Running total: 8 of ? biz days used
▶ Clock Running
Status: Scheduled
Apr 8 → Apr 14, 2026
4 business days (counted)
Running total: 12 of ? biz days used
Created: Jan 28, 2026Completed: Apr 14, 2026
Show SLA Clock Detail (5 periods)
Period
Status
Clock
Biz Days
Running Total
Jan 28 → Jan 28
Created
▶ Running
0
0
Jan 28 → Mar 27
Ready for Release
⏸ Paused
41
—
Mar 27 → Apr 8
Ready for Scheduling
▶ Running
8
8
Apr 8 → Apr 8
Assigned
▶ Running
0
8
Apr 8 → Apr 14
Scheduled
▶ Running
4
12
Dispatch0.0d
Holding42.0d (excluded)
Coordinators13.2d
Billing13.1d
✓ SLA Met
Dispatch
0.0 days
Holding(excluded from total)
42.0 days
Coordinators
13.2 days
Over SLA
Billing
13.1 days
Over SLA
Created: Jan 28, 2026
Due: Apr 27, 2026
Completed: Apr 14, 2026 (13 days early)
📅
53.4 daysTotal Age
⏱️
42.0d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
25.2d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
2 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 200 days old
🚨
Primary Delay42.0 days stuck in "Ready for Release"
⏱️
Total Delay25.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
42.0d
Ready for Scheduling
8.8d / 1.0d SLA
Assigned
0.0d / 0.5d SLA
Scheduled
4.4d / 2.0d SLA
Closed
4.0d / 1.0d SLA
Ready to Bill
13.0d / 1.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
04/07/2026 mklose rharris: Confirmed with Santiago Ramos, Radames Jorge radames.j.santiagoramos.civ@health.mil for Jordan to service 5 devices Monday 4/13...
03/27/2026 YEspino Recieved an email back from Radames confirmed the devices in question. He is ready to schedule asap...
03/25/2026: ARoosa - Sent fu email to Radames Santiago about devices in question, reminding him we spoke last week on the phone. Asked for confirmation of serial #'s & credentialing requirements. radames.j.santiagoramos.civ@health.mil...03/20/2026: ARoosa - Called (254) 553-1308, Radames Santiago,Med Equip Specialist - QA/QC Manager, He has been awaiting us to contact him and get these issues resolved. sending email to radames.j.santiagoramos.civ@health.mil so he can confirm serial #'s.
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01340256 - FCO - DD: 4/27/2026 - - Giraffe OmniBed Carestation - FORT HOOD - CARL R DARNALL ARMY MEDICAL CENTER
Apr 13, 2026 8:00am – 9:00am
Work Order: WO-01340256
Account: GE Healthcare FMI
Customer Case Number: 1-557093779071
Location: CARL R DARNALL ARMY MEDICAL CENTER
Address: 36065 SANTA FE AVE, FORT HOOD, TX 76544
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TABC70216
Contact Name: Radames Santiago,Me
PDF
TABC70216_Product Label
Apr 17, 2026
PDF
TABC70216_Screw 1
Apr 17, 2026
PDF
TABC70216_Screw 2
Apr 17, 2026
Timeline
📋
▶
2
Work Order Created Jan 28, 8:30 AM
Work Order Created → Status: Ready for Release
✅
Status: Ready for Release by Mandy Newman
From:Entered
To:Ready for Release
Duration in Previous:0m
🚨42.0 business days
📅
Status: Ready for Scheduling by Yessenia EspinoMar 27, 1:16 PM