Created: Feb 2, 2026
Due: Feb 5, 2027 (106 days left)
📅
180.8 daysTotal Age
⏱️
173.0d in Ready for ReleaseLongest Stage
🔄
4 transitionsStatus Changes
⚠️
8.0d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 261 days old
🚨
Primary Delay173.0 days stuck in "Ready for Release"
⏱️
Total Delay8.0 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
7.8d / 0.5d SLA
Ready for Release
173.0d
Scheduled
1.9d / 2.0d SLA
Closed
1.7d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
10/07/2026 MKlose-rcvd email from POC adv ship parts to location and wll do in house. looped in Fady .... 10/07/2026: ARoosa-Called 860-889-8331 xfer to biomed, line rang over and over so I disconnected. Called back attempting to reach operator to xfer to biomed instaed of automated xfer, line continuously rang as well until I disconnected. Found number for Ray Desocio, Called (860) 250-5649 to speak to Ray, spoke to Ray he said I would need David Doyle who handles the Carescape Monitors for Biomed gave me email to send info to at david.doyle@hhchealth.org, (860) 221-4709
🔧Work Performed
Template Description: FMI 36168 Code 1
Problem Found: FMI 36168 - Potential FLEX-3S2P battery failures with certain CARESCAPE
B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0
Patient Monitors
and Portrait VSM (MVC101)
Action Taken: FMI 36168 completed. Battery was replaced with part number 2062895-
001 Rev T from FMI36168 kit. Faulty battery was destroyed.
Communicated to the
customer to destroy all affected batteries including affected battery FRUs.
Verification Test: Verified that the replaced battery was affected.
When battery was
replaced for a monitor, verified that there was no “Battery failure“, “Condition battery“
or “Replace battery“ message displayed in the message field on the
monitor screen after
battery replacement. Equipment passed all required tests and has been returned to the
customer fully functional.
🔍Technician Findings
Template Description: FMI 36168 Code 1
Problem Found: FMI 36168 - Potential FLEX-3S2P battery failures with certain CARESCAPE
B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0
Patient Monitors
and Portrait VSM (MVC101)
Action Taken: FMI 36168 completed. Battery was replaced with part number 2062895-
001 Rev T from FMI36168 kit. Faulty battery was destroyed.
Communicated to the
customer to destroy all affected batteries including affected battery FRUs.
Verification Test: Verified that the replaced battery was affected.
When battery was
replaced for a monitor, verified that there was no “Battery failure“, “Condition battery“
or “Replace battery“ message displayed in the message field on the
monitor screen after
battery replacement. Equipment passed all required tests and has been returned to the
customer fully functional.
⚠️Problem Description
FMI 36168: Potential FLEX-3S2P battery failures with certain CARESCAPE B450, CARESCAPE ONE,
B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait VSM (MVC101)
Purpose: Replace and destroy/scrap affected batteries delivered with CARESCAPE B450, CARESCAPE
ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait VSM (MVC101)
monitors, and separate affected FRU batteries. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
Patient Monitor
📞Call Description
FMI 36168
📅
WO-01341668 - FCO - DD: 2/5/2027 - - Carescape One Monitor - FRANKLIN - BACKUS HOSPITAL WAREHOUSE
Oct 20, 2026 8:00am – 5:00pm
Work Order: WO-01341668
Account: GE Healthcare FMI
Customer Case Number: 1-556825775053
Location: BACKUS HOSPITAL WAREHOUSE
Address: 11 MURPHY RD, FRANKLIN, CT 06254
Manufacturer: GE
Model #: Carescape One Monitor
Serial #: SV925060286SA
Contact Name: David Doyle
Contact Phone: (860) 221-4709
Contac
UNKNOWN
SV925060286SA_Product label pic
Oct 20, 2026
Timeline
📋
Work Order Created Feb 2, 12:28 PM
Work Order:WO-01341668
Type:FCO
Priority:PC
⚠️8.0 business days
✅
Status: Ready for Release by Mandy NewmanFeb 11, 9:51 AM