📍 PENNSYLVANIA HOSPITAL — 800 SPRUCE ST, PHILADELPHIA PA, 19107
Feb 12, 2026 → Jul 1, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Mar 28, 2026
✗ SLA Missed (66 biz days late)
Active: 96 biz days
🕐 Clock stopped: Canceled (Jul 1, 2026)
17d
▶ Clock Running
Status: Created
Feb 12 → Mar 10, 2026
17 business days (counted)
Running total: 17 of ? biz days used
79d
▶ Clock Running
Status: On Hold
Mar 10 → Jul 1, 2026
79 business days (counted)
Running total: 96 of ? biz days used
Created: Feb 12, 2026Canceled: Jul 1, 2026
Show SLA Clock Detail (2 periods)
Period
Status
Clock
Biz Days
Running Total
Feb 12 → Mar 10
Created
▶ Running
17
17
Mar 10 → Jul 1
On Hold
▶ Running
79
96
Dispatch18.0d
Holding80.0d (excluded)
✗ SLA Missed
Dispatch
18.0 days
Over SLA
Holding(excluded from total)
80.0 days
Created: Feb 12, 2026
Due: Mar 28, 2026
Completed: Jul 1, 2026 (94 days late)
📅
97.0 daysTotal Age
⏱️
80.0d in On HoldLongest Stage
🔄
2 transitionsStatus Changes
⚠️
17.5d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
0 linesWork Details
AI Analysis
Powered by OpenAI
Generating AI analysis...
Behind Schedule - Needs Attention
Canceled • 183 days old
🚨
Primary Delay80.0 days stuck in "On Hold"
⏱️
Total Delay17.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
18.0d / 0.5d SLA
On Hold
80.0d
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/12/2026 SRobinson- Created follow up WO as requested. (Old WO-01282603) -Follow up notes - 2-09-26 Michael Steencken “Completed software upgrade only. Need on-site follow-up WO to return and replace batteries when they become available”
-Previous notes -
02/04/2026 RFosbury Sending FSE Mike S onsite Feb 9th
01/06/2026 Rfosbury called (267) 357-3819, call went to VM/LM
09/22/2025 mklose called (215) 316-5151, xfer to operator, no one answered
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
Timeline
📋
Work Order Created Feb 12, 10:47 AM
Work Order:WO-01351315
Type:FCO
Priority:PC
🚨18.0 business days
⏸️
Status: On Hold by Mandy NewmanMar 10, 9:28 AM
From:Entered
To:On Hold
Duration in Previous:25d 21h
🚨80.0 business days
🚫
Status: Canceled by Mandy NewmanJul 1, 8:25 AM
From:On Hold
To:Canceled
Duration in Previous:112d 22h
Bottlenecks
Entered
Duration: 18.0 business days (threshold: 4.0 business hours)