Work Order 360 View
Detailed work order timeline, KPIs, and analysis
P5
5 business days from creation
Target:
Feb 24, 2026
🕐 Clock stopped: Completed (Feb 18)
✓ SLA Met (5d early)
Coordinators
1.8 days
Over SLA
Created: Feb 17, 2026
Due: Feb 24, 2026
Completed: Feb 18, 2026 (6 days early)
⏱️
7.4d in Closed
Longest Stage
🔄
5 transitions
Status Changes
📦
0 orders, 0 quotes
Parts & Quotes
AI Analysis
Powered by OpenAI
Generating AI analysis...
🚨
Primary Delay
7.4 days stuck in "Closed"
⏱️
Total Delay
7.7 business days over SLA thresholds
SLA vs Actual Time by Status
Ready for Scheduling
0.0d
/ 1.0d SLA
Ready to Bill
0.1d
/ 1.0d SLA
💬 Activities, Notes & Communications
🔧
Work Performed
replaced main control board and footboard connector cable, wiped down bed and tested all functions and returned bed to service
🔍
Technician Findings
main board had corrosion on footboard connector cable pins
⚠️
Problem Description
repair request
no function on footboard
Lancara, Kenneth
📄
Description
FL28EX GoBed II
Timeline
📋
📅
From:
Entered
To:
Ready for Scheduling
Duration in Previous:
0m
👷
👷
From:
Ready for Scheduling
To:
Assigned
Duration in Previous:
7m
⏱️
✔️
From:
Assigned, Completed
To:
Completed, Closed
Duration in Previous:
1d 2h, 0m
📄
From:
Closed
To:
Ready to Bill
Duration in Previous:
8d 18h
📨
From:
Ready to Bill, Invoice Pending
To:
Invoice Pending, Invoiced
Duration in Previous:
42m, 0m
🧾
Invoice Number:
SM-255180
Bottlenecks
Assigned
Duration: 1.8 business days (threshold: 4.0 business hours)
Exceeded by 1.3 business days
Closed
Duration: 7.4 business days (threshold: 1.0 business days)
Exceeded by 6.4 business days
ℹ️ Work Order Details
Technician
Kenneth Lancara
Work Details (4)
| Name |
Type |
Status |
Qty |
Amount |
Date |
| WL-03110690 |
Labor |
Open |
0.25 |
$0.00 |
Feb 18, 3:27 PM |
| WL-03110691 |
Labor |
Open |
2.5 |
$0.00 |
Feb 18, 3:27 PM |
| WL-03110692 |
Parts |
Open |
1 |
$0.00 |
Feb 18, 3:27 PM |
| WL-03110693 |
Parts |
Open |
1 |
$0.00 |
Feb 18, 3:27 PM |