The SpO2 monitor is not working at
all. We have tried switching out the sensors and line and neither is working. The red light in the sensor is on and blinking but the monitor states the
sensor is off
Model: Smith Medical - SurgiVet Advisor tech, V9219
Total Delay162.8 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Awaiting Parts
163.9d / 3.0d SLA
On Hold
14.6d
Tech On Site
0.5d / 1.0d SLA
Incomplete
3.8d
Assigned
1.8d / 0.5d SLA
Completed
0.0d / 1.0d SLA
Closed
3.7d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
4/23-RS; item is available through Patterson Vet, who will not sell to service companies. Every procurement channel we tried either was not available or our order was canceled due to being unable to fulfill. MM cannot action this further, please request cancellation and ask the customer to order it directly
.12-22 DSR Sent email to Depot Q Customer requesting Update on ETA.
9/9 KHardison, Added quote approved quote, Q-00029943, for parts
Visual inspection; Passed
Power Outlet Test; Passed
Power cord and plug; Passed
Functional Inspection; Passed
ECG measurement (± 5 bpm)
Normal: 80bpm, 1mV = __80__ bpm
Leads Off Detection; Passed
SpO2 Measurements:
90% (±2%) = __88___%
NIBP Measurements:
0 mmHg (±1 mmHg) = __0___mmHg
200 mmHg (±1 mmHg) = __199___mmHg
Temperature measurement(± 0.1 C)
T1: 37.0C = __37.0__ C; Passed
T2: 37.0C = __37.0__ C; Passed
NIBP Leak Test; Passed
Loudspeaker; verified
Cleaned device externals
Will need new SpO2 sensor to go back into service in good working condition. Unable to order sensor and sensor will need to be purchased on customer's end.
Depot Shipping Box - 24 x 18 x 12" Corrugated Boxes
Shipment Tracking Number: 1Z01E3E70393821155
- Estimated delivery : 06/26/2025
RMA Tracking Number: 1Z01E3E79094403763.
_____________________________
item is available through Patterson Vet, who will not sell to service companies. Customer must order replacement sensor through a their purchasing group.
🔍Technician Findings
Unit was received into our Depot. Was able to duplicate issue with customer's sensor cable. Was not able to duplicate with our sensor cable. Unit will need new SpO2 sensor to fix issue.
Please quote for:
1 x *NEW PART* PN: V3078 (Oximetry Sensor Mini Clip)
https://mms.mckesson.com/product/936496/ICU-Medical-FKA-Smiths-Medical-V3078
2 x Labor (Repair,PM)
3 x Shipping
4/23/26 SKLEIN - Unable to purchase part and part will need to be purchased by customer.
⚠️Problem Description
Q-00029943, parts.
The SpO2 monitor is not working at
all. We have tried switching out the sensors and line and neither is working. The red light in the sensor is on and blinking but the monitor states the
sensor is off
Model: Smith Medical - SurgiVet Advisor tech, V9219 https://www.ebiotrack.com/workorder.php?wo=67572
📧
Email: EMSAR WO-01298454
Completed Jan 8, 2026
To: kaydie.ahern@lovet.com
CC:
BCC: ryan.schwochow@emsar.com
Attachment: --none--
Subject: EMSAR WO-01298454
Body:
I hope you're doing well. I wanted to provide you with an update regarding your service request for WO-01298454. We are currently waiting to receive tracking information for the parts needed to complete the [repair/maintenance].
Once the tracking details are confirmed and the parts are on their way, I will reach out to let you know next steps.
Thank you for your patience, and
📧
Email:Re: Customer Requesting Update on WO-01298454
Completed Jan 8, 2026
From: derek.reynolds@emsar.com
To: depotsupport@emsar.com
Cc: ryan.schwochow@emsar.com, samuel.klein@emsar.com
Work Order
WO-01298454
Derek S. ReynoldsVirtual Tech Teamderek.reynolds@emsar.com888-841-5155 ext 2Office: (937) 741-4833EMSAR Confidentiality Notice: This electronic mail message, including any attachments, may contain information that is privileged, confidential, proprietary in nature, or otherwise private information that is protected by law from disclosure or subject to copyright
📋
Inbound from +16233990141
Completed Dec 22, 2025
Transferred to Derek
📧
Email: Reminder: Send Your Device for Depot Repair
Completed Sep 2, 2025
To: kaydie.ahern@lovet.com
CC:
BCC:
Attachment: --none--
Subject: Reminder: Send Your Device for Depot Repair
Body:
Dear Customer, Thank you for choosing EMSAR for your Depot Repair. We have initiated a repair work order for your device. However, we haven't received it at our depot yet. Work Order #: WO-01298454 Location: Lovet Pet Health Care - Happy Valley Animal Hospital Address: 6615 W. Happy Valley Rd., Suite B106 Glendale, AZ 85310 Device: The SpO2
📧
Email: Reminder: Send Your Device for Depot Repair
Completed Aug 26, 2025
To: kaydie.ahern@lovet.com
CC:
BCC:
Attachment: --none--
Subject: Reminder: Send Your Device for Depot Repair
Body:
Dear Customer, Thank you for choosing EMSAR for your Depot Repair. We have initiated a repair work order for your device. However, we haven't received it at our depot yet. Work Order #: WO-01298454 Location: Lovet Pet Health Care - Happy Valley Animal Hospital Address: 6615 W. Happy Valley Rd., Suite B106 Glendale, AZ 85310 Device: The SpO2
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Aug 19, 2025
To: kaydie.ahern@lovet.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01298454Location: Lovet Pet Health Care - Happy Valley Animal HospitalAddress: 6615 W. Happy Valley Rd., Suite B106 Glendale, AZ 85310Device: Problem Description: The SpO2 monitor is not
Oct 6, 2025 12:09pm
@Ryan Schwochow
Do we know what we're pending on this one?
RMA tracking shows delivered: 1Z01E3E79094403763
Apr 23, 2026 8:11am
@Samuel Klein We cannot buy the sensor. Labor only and send back
PDF
ShippingLabel_0001_1Z01E3E79094403763.PDF
Aug 21, 2025
Timeline
📞
Case Linked Aug 19, 12:56 PM
Case Number:00520280
Subject:<no subject>
Status:Closed
📋
▶
3
Work Order Created Aug 19, 1:28 PM
Work Order Created → Parts Requested → Status: Awaiting Parts
📦
Parts Requested
Order:00080248
Status:Closed
📦
Status: Awaiting Parts by Derek Reynolds
From:Entered
To:Awaiting Parts
Duration in Previous:1m
⏳2.0 business days
💵
▶
3
3x Expenses Added Aug 20, 3:04 PM
3x Expenses Added → Parts Requested → Status: On Hold
📦
Parts Requested
Order:00080358
Status:Closed
⏸️
Status: On Hold by Brittany Peters
From:Awaiting Parts
To:On Hold
Duration in Previous:1d 1h
🚨12.0 business days
👷
▶
2
Technician Assigned Sep 4, 10:21 AM
Technician Assigned → Status: Assigned
👷
Status: Assigned by Samuel Klein
From:On Hold, Assigned
To:Assigned, Tech On Site
Duration in Previous:14d 19h, 0m
⚠️
Status: Incomplete by Samuel KleinSep 4, 1:15 PM
From:Tech On Site
To:Incomplete
Duration in Previous:2h 54m
⏳4.0 business days
👷
▶
2
Status: Assigned Sep 9, 3:18 PM
Status: Assigned → Technician Assigned
👷
Technician Assigned by Kimberly Hardison
Technician:Samuel A Klein
📦
▶
2
Parts Requested Sep 10, 8:29 AM
Parts Requested → Status: Awaiting Parts
📦
Status: Awaiting Parts by Ryan Schwochow
From:Assigned
To:Awaiting Parts
Duration in Previous:17h 10m
🚨162.0 business days
👷
Status: Assigned by Ryan SchwochowApr 23, 8:08 AM
From:Awaiting Parts
To:Assigned
Duration in Previous:224d 23h
⏱️
▶
2
Labor Added Apr 23, 9:44 AM
Labor Added → Status: Tech On Site
🔧
Status: Tech On Site by Samuel Klein
From:Assigned
To:Tech On Site
Duration in Previous:1h 36m
✔️
▶
3
Status: Completed Apr 23, 10:30 AM
Status: Completed → Parts Requested → Status: Awaiting Parts
📦
Parts Requested
Order:00095060
Status:Closed
📦
Status: Awaiting Parts by Samuel Klein
From:Completed
To:Awaiting Parts
Duration in Previous:1m
💵
▶
2
Expenses Added Apr 23, 10:56 AM
Expenses Added → Status: On Hold
⏸️
Status: On Hold by Sean Weaver
From:Awaiting Parts
To:On Hold
Duration in Previous:24m
⏳3.0 business days
🏁
Status: Closed by Ryan SchwochowApr 27, 2:32 PM
From:On Hold
To:Closed
Duration in Previous:4d 3h
⏳4.0 business days
📄
▶
2
Status: Ready to Bill Apr 30, 11:11 AM
Status: Ready to Bill → Invoice Created
🧾
Invoice Created
Invoice Number:SM-260518
Bottlenecks
Assigned
Duration: 1.6 business days (threshold: 4.0 business hours)
Exceeded by 1.1 business days
Awaiting Parts
Duration: 162.0 business days (threshold: 3.0 business days)
Exceeded by 159.0 business days
Closed
Duration: 3.7 business days (threshold: 1.0 business days)