14W
SYS-LC-1260II
Contract ends 01/31/2026 includes Repair
ENHPM / EOQ
Enhance PM -, the lamp is automatically changed at time of PM service with no questions. Have techs just order a lamp for an Enhanced PM.
G7111B
DEAET03460 1260 Infinity II Quaternary Pump
G7116A
DEAEM09362 1260 Infinity II Multicolumn Thermostat
G7114A
DEACX26188 1260 Infinity II VW Detector
G7129A
DEAEQ47099 1260 Infinity II Vialsampler
12/11 Part still can't be added, WO will be closed is SM and SO will be closed in mService. MD
12/09/2025 Google map was used to calculate zone. An email was sent to add parts G1314-67188 in SM. A repair SO 6008276703. was opened. MD
Please assign to Maria Davis ARP to site is 42 miles. Zone 1
🔧Work Performed
12/09/2025 ACE Enhance PM session was created. Pm was completed per Agilent. System was flushed thoroughly with H2O. New lamp was installed and calibrated, stabilized for some time. Intensity, dark current, filter, cell test but Holmium passed.
⚠️Problem Description
14W
SYS-LC-1260II
Contract ends 01/31/2026 includes Repair
ENHPM / EOQ
Enhance PM -, the lamp is automatically changed at time of PM service with no questions. Have techs just order a lamp for an Enhanced PM.
G7111B
DEAET03460 1260 Infinity II Quaternary Pump
G7116A
DEAEM09362 1260 Infinity II Multicolumn Thermostat
G7114A
DEACX26188 1260 Infinity II VW Detector
G7129A
DEAEQ47099 1260 Infinity II Vialsampler