11/07/2025 EMartin dispatching to FSE onsite per request for SN# H32900195BE66
located at MedOne in Raleigh, NC
🔧Work Performed
H32900195BE66 - System PCBA replacement needed. Pilot sensor zero (0) reading during testing. Multiple sensor errors. Software 1.05.10.00 reloaded. System placed in ship mode. Three-way and proportional valves replaced. Device reassembled and performed a full diagnostic test. DUT continues to fail pilot [System PCBA sensor] reading. Incident case open, and system PCBA needs to be added to quote for approval.
WO-01313785 - FCO - Trilogy EV300 Vent - Raleigh - Med One Group - Raleigh