11/14 KKelch Tim called me back to reschedule for 11/19 between 12 and 2 pm. Customer confirmed.
11/13 KKelch called Tim to reschedule due to routing needs for 11/19 between 12 and 2 pm. Left VM.
11/12 KKelch called Tim to schedule for 11/17 between 12 and 2 pm. Customer confirmed.
🔧Work Performed
I removed the legs and replaced the 16 half shells I then removed the sensor housings and replaced the crest to crest springs slider and slider magnets I then replaced the Left and right slider housings I installed the two new lift here labels and installed three new zip ties I then replaced the missing nylon washer nut and washer I then completed the pm on the unit I function tested the unit and all passed I returned the cot to the service.
🔍Technician Findings
Disinfect and clean cot mattress restraints and
accessories inspected cot to confirm the
product is free from pinch or cut hazards All
welds and assemblies are intact hardware and
wires are properly securedbr brCycle tested cot in
all positions in both power and manual backup
modebr brLoaded and unloaded cot with fastener
in both power and manual modebr brCot fully
operational reviewed inspection results with
customer returned cot to servicebr On the 6506 I
arrived and found that the cot was struggling to
raise The bearings were worn The slider and
slider magnet were damaged and the cresttocrest
springs were broken The left and right Slider
housings were grooved and needing to be
replaced The upper and lower Xframe guards
were damaged and needed to be replaced I then
discovered that the cot was missing a nylon
washer and was missing a nut and washer on the
foot end fastener assembly I also discovered that
the left and right Lift here labels were torn the
three zip ties were missing