***PART RETURN IS REQUIRED IF RMA IS NOT IN BOX, Email ferno@emsar.com for label***
STRETCHER IS MOVING SLOW AND HAS 6 WRENCH FLASHES. CHANGE ACTUATOR AND MAIN HARNESS. RETURN ALL PARTS
Part ID: 0806119
Return Parts: Yes
KITACTUATOR REFURB POWERX1 FH
Part ID: 2183114
Return Parts: Yes
WIRE HARNESS
11/12/25 SReich - Left Michael a vm, gave him the WO # and asked that he let us know when parts are received.
🔧Work Performed
On the X1, I removed the actuator and disconnected all the wires and zip ties. I then installed the new actuator provided and connected all the wires and zip ties. I then tested the cot, and the cot was still showing 6 wrenches and raising slowly. I then put the new main harness on the outside of the cot and tested the cot. The cot was still raising slowly. I then found that the LVDT harness had a short in the harness. I pulled a new LWDT harness from my stock and installed it on the cot. The cot then started the raise normally. I reconnected the original main harness and I left the new hydraulic assy. in the cot. I function tested the cot and all passed. I returned the cot to service.
🔍Technician Findings
On the X1, I arrived and found that the cot was raising very slow and the cot was showing 6 wrench flashes.
⚠️Problem Description
***PART RETURN IS REQUIRED IF RMA IS NOT IN BOX, Email ferno@emsar.com for label***
STRETCHER IS MOVING SLOW AND HAS 6 WRENCH FLASHES. CHANGE ACTUATOR AND MAIN HARNESS. RETURN ALL PARTS
Part ID: 0806119
Return Parts: Yes
KITACTUATOR REFURB POWERX1 FH
Part ID: 2183114
Return Parts: Yes
WIRE HARNESS
📄Description
Ferno Power X1 Power Cot
📋
Outbound to +19729225638
Completed Nov 12, 2025
Left Michael a vm, gave him the WO # and asked that he let us know when parts are received.
Nov 19, 2025 12:59pm
11/19 KWeiss, Reviewed pictures and closing w/o for billing review.
📅
WO-01314745 - Repair - Power X1 - Carrollton - ALLEGIANCE MOBILE HEALTH (XA Customer)
Nov 17, 2025 4:00pm – 6:00pm
Work Order: WO-01314745
Account: FERNO
Customer Case Number: PCS2501918 Claim #752
Location: ALLEGIANCE MOBILE HEALTH (XA Customer)
Address: 1681 S Broadway, Carrollton, TX 75006
Manufacturer: Ferno
Model #: Power X1
Serial #: 24C209956
Contact Name: Michael Gilbert
Contact Phone: 9729225638
Contact