Primary Delay4.0 days stuck in "Ready for Release"
⏱️
Total Delay4.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
1.2d / 0.5d SLA
Ready for Release
4.0d
Scheduled
0.4d / 2.0d SLA
Tech On Site
0.1d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
5.2d / 1.0d SLA
WO Correction Needed
1.2d
07e-Partial Project
3.0d
Ready to Bill
1.6d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
11/26/2025 SRobinson- Created follow up WO as requested. (WO-01317268)
Quote accepted
11/25/2025 EMartin dispatching to FSE per request
11/24/2025 EMartin per Fady working with Chad on these repairs will provide resource once completed
11/20/2025 EMartin Fady working with Chad on remaining devices
🔧Work Performed
Three-way and proportional valves replaced. System continues to fail FCO pre-checks. System passed all diagnostic tests prior to pneumatic test. Pressure Verification: Setpoint 80: Pilot: Difference -79.88 cmH2O -2.00 2.00 FAIL. System PCBA replacement required.
🔍Technician Findings
Parts needed.
⚠️Problem Description
FCO86600081. Repair to be handled by Philips ASP (CSA/EMSAR
WO-01316421 - Repair - Trilogy EV300 Vent - Raleigh - Med One Group - Raleigh
Nov 25, 2025 8:00am – 5:00pm
Work Order: WO-01316421
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-13942260
Location: Med One Group - Raleigh
Address: 3811 Tarheel Drive Suite 109, Raleigh, NC 27609
Manufacturer: Philips Healthcare
Model #: Trilogy EV300 Vent
Serial #: H32900195BE66
Contact Name: Chad Agliam
Contact Phon