FMI 78101C: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.4 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
11/21/2025 Cever dispatched to FSE per their request WO needed
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R3.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass.Completed all Peripheral Checks-system passed.Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
Good working conditions
⚠️Problem Description
FMI 78101C: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.4 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📅
WO-01316577 - FCO - Venue Fit Ultrasound - EASTON - ST LUKES HOSPITAL ANDERSON
Nov 21, 2025 8:00am – 5:00pm
Work Order: WO-01316577
Account: GE Healthcare FMI
Customer Case Number: 1-554880082257, 1-554907752088
Location: ST LUKES HOSPITAL ANDERSON
Address: 1872 ST LUKES BLVD, EASTON, PA 18045
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFC000753
Contact Name: Matthew Burian
Contac
JPG
VFC000753_New Battery
Nov 22, 2025
JPG
VFC000753_Diagnostic test report
Nov 22, 2025
JPG
VFC000753_Old Battery
Nov 22, 2025
JPG
VFC000753_Venue Fit
Nov 22, 2025
Timeline
📞
Case Linked Nov 19, 5:11 PM
Case Number:00545413
Subject:FMI 78101 FU WO Request
Status:Closed
📋
Work Order Created Nov 20, 3:02 PM
Work Order:WO-01316577
Type:FCO
Priority:PC
📅
Status: Ready for Scheduling by Sherry ReichNov 20, 3:08 PM