FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
SN TAB24180005SA
Primary Delay43.3 days stuck in "Ready for Scheduling"
⏱️
Total Delay86.7 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
66.2d / 1.0d SLA
Ready for Release
17.4d
Scheduled
18.1d / 2.0d SLA
Tech On Site
0.1d / 1.0d SLA
Closed
4.0d / 1.0d SLA
Ready to Bill
19.7d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
04/10/2026 - ARuby - Dispatching to Jeffrey Coles for 4/13. Sent confirmation email to Jody and Kevin Glover Director KRGLOVER@salisbury.edu (cell) (215) 820-2695 (Office) (410) 546-2465 (Jody is POC, if you can't reach her try Kevin).
04/09/2026 - ARuby - Emailed jwdengler@salisbury.edu Informing Jody of due date and proposed an earlier date. Put WO back in RFS
04/08/2026 - ARuby - Received confirmation email from Jody approving reschedule date. Dispatching to Jeffrey Coles 4/24
04/08/2026 - ARuby - called Jody (410) 546-2465 to reschedule for 4/24 VM/LM FU email to jwdengler@salisbury.edu
04/07/2026 - ARuby - Reassigning to Rick Charles 4/17
04/07/2026 - ARuby - called Jody Dengler (410) 546-2465 jwdengler@salisbury.edu Confirmed date dispatching FSE Jeffrey Coles 4/17
02/04/2026 RFosbury working to reschedule
02/03/2026 RFosbury called to reschedule. confirmed FSE onsite Feb 6th,
01/20/2026 RFosbury Called Jody Dengler and confirmed FSE onsite Jan 30th
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
FMI32097
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
SN TAB24180005SA
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📧
Live Call_Followed Up With Email
Completed Apr 8, 2026
04/08/2026 - ARuby - called Jody (410) 546-2465 to reschedule VM/LM FU email to jwdengler@salisbury.edu
📧
Live Call_Followed Up With Email
Completed Apr 7, 2026
04/07/2026 - ARuby - called Jody Dengler (410) 546-2465 jwdengler@salisbury.edu Confirmed date dispatching FSE Jeffrey Coles 4/17
📧
Email: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - SALISBURY UNIVERSITY CENTRAL RECEIVING - SALISBURY, MD - WO-01316586
Completed Jan 20, 2026
To: jwdengler@salisbury.edu
CC: jeffrey.coles@emsar.com; douglas.pearce@emsar.com; rich.fosbury@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: FMI32097 MIC UMDC English Customer Letter.pdf
Subject: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - SALISBURY UNIVERSITY CENTRAL RECEIVING - SALISBURY, MD - WO-01316586
Body:
Hi Jody,
Thank you for taking my call today. As discussed, I am contacting you on behalf of GE Healthcare about a Field Safety N
📧
Live Call_Followed Up With Email
Completed Jan 20, 2026
01/20/2026 RFosbury Called Jody Dengler and confirmed FSE onsite Jan 30th
📅
WO-01316586 - FCO - DD: 4/27/2026 - - Giraffe OmniBed Carestation - SALISBURY - SALISBURY UNIVERSITY CENTRAL RECEIVING
Apr 13, 2026 8:00am – 5:00pm
Work Order: WO-01316586
Account: GE Healthcare FMI
Customer Case Number: 1-556536645153
Location: SALISBURY UNIVERSITY CENTRAL RECEIVING
Address: 106 PINE BLUFF RD, SALISBURY, MD 21804
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24180005SA
Contact Name: Jody Dengler
JPG
SCREW 1
Apr 13, 2026
JPG
GOB_TAB2418005SA
Apr 13, 2026
JPG
SCREW 2
Apr 13, 2026
Timeline
📞
Case Linked Nov 19, 5:24 PM
Case Number:00545417
Subject:11/20 waiting on response from Ally
Status:Closed
📋
▶
2
Work Order Created Nov 20, 3:26 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Sherry Reich
From:Entered
To:Ready for Scheduling
Duration in Previous:3m
🚨22.0 business days
✅
Status: Ready for Release by Erin MartinDec 23, 3:28 PM