12/3 I have evaluated this unit and have found it only raises a little bit at a time. I have tested the buttons and the hall effects sensor. I have found the Hall sensor is not reading the magnet. I have removed and replaced the hall effects sensor and the sensor wire due to corrosion on both. Functional testing passed. This unit is ready to return to service
🔍Technician Findings
Unit raises a little bit at a time.
⚠️Problem Description
Q-00031411
Stryker cot SN 08141121 needs the following part for repairs.
Chane is onsite today 12/3 with parts.
PN#6500-001-161 X1
PN# 6500-001-160 X1 https://www.ebiotrack.com/workorder.php?wo=70113
📄Description
6500 Power Pro IT
Timeline
📞
Case Linked Dec 3, 4:13 PM
Case Number:00548610
Subject:TECH ONSITE - WO NEEDED
Status:Closed
📋
▶
4
Work Order Created Dec 4, 9:17 AM
Work Order Created → Status: Ready for Scheduling → Technician Assigned → Status: Assigned
📅
Status: Ready for Scheduling by Sherry Reich
From:Entered
To:Ready for Scheduling
Duration in Previous:1m
👷
Technician Assigned by Kaitlyn Kelch
Technician:Chane Lacy
👷
Status: Assigned by Kaitlyn Kelch
From:Ready for Scheduling
To:Assigned
Duration in Previous:1m
🔩
3x Parts Added Dec 4, 7:23 PM
Type:Parts, Labor
Status:Open
Qty:1
✔️
Status: Completed by Chane LacyDec 4, 7:38 PM
From:Assigned, Completed
To:Completed, Closed
Duration in Previous:10h 18m, 0m
⏳3.5 business days
📄
▶
3
Status: Ready to Bill Dec 9, 11:29 AM
Status: Ready to Bill → Invoice Created → Status: Invoiced
🧾
Invoice Created
Invoice Number:SM-248541
💰
Status: Invoiced by Diane Patton
From:Invoice Pending
To:Invoiced
Duration in Previous:1m
Bottlenecks
Assigned
Duration: 1.1 business days (threshold: 4.0 business hours)
Exceeded by 4.6 business hours
Closed
Duration: 3.0 business days (threshold: 1.0 business days)