Midmark Service Request Dispatch Authorization - WO-00139560
Closed
Zone 1
Dispatch Type Repair Email info@karterplasticsurgery.com
PROBLEM SUMMARY
WD Unusual noise
ADDITIONAL NOTES
Closes at 3pm on Fridays. For assistance, troubleshooting, or parts identification, contact Medical Tech Service at
844-856-1230. Thank you!
12/22 KSmith -- Scheduled for 12/29 per Sarah's request. 8-10am. Site is closed 12/26.
12/19 KHardison - Left a vm regarding WO. Advised that a coordinator will call for scheduling,
🔧Work Performed
**Upon eval and speaking with staff they showed me a video of their complaint. I ran an unwrapped cycle and opened door when prompted and unit ran full cycle with no issues. Spoke with Brian @MM and he confirmed that if door isn’t opened when prompted it can damage unit and cause funny noises. Spoke with Sienna (office manager) she let me know that they used to have an M9 that opened on its own. Unit is in another room in the back and they dont know when door needs to be opened. She informed me that she will be contacting her rep to get an autoclave that the door opens on its own. I assisted in getting her any info i could to facilitate.
🔍Technician Findings
Unit in service
⚠️Problem Description
Zone 1
Dispatch Type Repair Email info@karterplasticsurgery.com
PROBLEM SUMMARY
WD Unusual noise
ADDITIONAL NOTES
Closes at 3pm on Fridays. For assistance, troubleshooting, or parts identification, contact Medical Tech Service at
844-856-1230. Thank you!
Work Order: WO-01327392
Account: MIDMARK
Customer Case Number: SR 1511404 WO-00139560
Location: Karter Advanced Facial Plastic Surgery
Address: 120 Hebron Ave, Ste 2A,, Glastonbury,, CT 06033-2096
Manufacturer: Midmark
Model #: M9 (Series) Sterilizer
Serial #: V2846145
Contact Name: Sarah Bell
Conta