Nkosi Wilson - Upon initial inspection of device, DUT failed AECM Verification Testm 12/17/2025. Informed Trevor Edens of repair requirements and that FCO may not be implemented until repair is complete. Customer Request Philips ASP to perform repair.
(DO NOT ORDER PARTS).
Valve, proportional (AECM) 453561539901 Qty. 1
Valve, three-way solenoid 453561539911 Qty. 1
04/27/2026 RArnold - Michael completed FCO on 04/17. Repair not needed. Canceling this EWO
04/20/2026 - ARuby - Asked michael if he's received PO for WO so we can create new PWO.
04/15/2026 - ARuby - Assigning to Michael per Doug. New PWOs need created
04/02/2026 DMalloy - Reassigning WO to Rodney per Fadys request.
03/05/2026 SRobinson- Reassigning WO to Rodney per Fadys request.
02/17/2026 RArnold - PWO Canceled. DO NOT cancel EWO so we do not lose visibility
12/31/2025 RArnold - PWO pending entitlement
12/19/2025 RArnold -WO for device that failed FCO81. PWO pending entitlement
⚠️Problem Description
Nkosi Wilson - Upon initial inspection of device, DUT failed AECM Verification Testm 12/17/2025. Informed Trevor Edens of repair requirements and that FCO may not be implemented until repair is complete. Customer Request Philips ASP to perform repair.
(DO NOT ORDER PARTS).
Valve, proportional (AECM) 453561539901 Qty. 1
Valve, three-way solenoid 453561539911 Qty. 1