Midmark Service Request Dispatch Authorization - WO-00140060
Closed
Zone 2
WD - Arm drifting - Run with 1511866
Imaging/X-Ray Dispatch. This dispatch needs to be completed on the same day as SR 1511866. Will need set of hex keys as
well as 9 inch 8 mm t-allen wrench. For assistance, troubleshooting, or parts identification, contact Veterinary Tech Service at
844-856-1232, option 3. Thank you!
All x-ray WOs:
Mandatory documents:
FDA 2579 Form
12/22 MOwens Called and spoke to Kelly Lucchesi. She asked what would be the first available opening for the visit for he tech. I told her next Tues. 12/30 in the afternoon. She said that day will work for her. Arrival will be around 3pm.
12/22/2025jsells
spoke with Kelly, and I gave her the work order number and told her a coordinator will call to schedule
🔧Work Performed
an 8mm allen wrench used to adjust the drift for the tube head. the unit was set to one position and no movement was observed.
🔍Technician Findings
the tube head drifts up after being set.
⚠️Problem Description
Zone 2
WD - Arm drifting - Run with 1511866
Imaging/X-Ray Dispatch. This dispatch needs to be completed on the same day as SR 1511866. Will need set of hex keys as
well as 9 inch 8 mm t-allen wrench. For assistance, troubleshooting, or parts identification, contact Veterinary Tech Service at
844-856-1232, option 3. Thank you!
All x-ray WOs:
Mandatory documents:
FDA 2579 Form