Oak Street Health - 24 Hour - 135889: New Work Order Assigned
Closed
Work Order #135889
Work Order Description:
Lab fridge in need or diagnosis and recalibration asap.
I contacted out IT for help with the probes for the dickson one thermometer and they look to be fine. The thermometer has been beeping alerting of out or range temperature but our records show that it is within range.
Assisted troubleshooting fridge Data logger issues.
1/6 emailed site for additional info. Maria is going to call VT when avail to troubleshoot.
🔍Technician Findings
Dickson 36 & Fridge is 37.4/3C. Setting up PID and changing to F for read out. Probe is not fully filled with Glycol, Sending Glycol to clinic to fill and had them move probe next to fridge probe.
American Biotech Supply - PH-ABT-HC-UCBI-0404G-ADA-LH
Model Type: Refrigerator
Asset ID: 12926224
Serial #: ABS-7204191-1806-V52
⚠️Problem Description
Work Order #135889
Work Order Description:
Lab fridge in need or diagnosis and recalibration asap.
I contacted out IT for help with the probes for the dickson one thermometer and they look to be fine. The thermometer has been beeping alerting of out or range temperature but our records show that it is within range. https://www.ebiotrack.com/workorder.php?wo=70689
SNOTE
VT Troubleshooting Notes
Jan 6, 2026
SNOTE
VT Notes
Jan 6, 2026
Timeline
📞
Case Linked Jan 6, 12:04 PM
Case Number:00556405
Subject:Oak Street Health - 24 Hour - 135889: New Work Order Assigned