Unit failed still failing after replacing 3 Way Solenoid Valve & Proportional Valve (AECM) , FCO86600081
unable to create subcase on Case # 0124915938 - please reference WO-13695670
Troubleshot the device and found that the tubing from the proximal port to the pressure
sensor on the system board was not securely attached, causing a leak and a failed pressure
test. The tubing has now been firmly re-attached, then Performed a complete PVT as per the
MFG's specifications. All tests and calibrations were found to be in spec or passed tests.
Device placed back into full clinical service; ready for patient use.
🔍Technician Findings
Repair.
⚠️Problem Description
Unit failed still failing after replacing 3 Way Solenoid Valve & Proportional Valve (AECM) , FCO86600081
unable to create subcase on Case # 0124915938 - please reference WO-13695670