eBiotrack - New Work Order Request - Inspection Repair
Closed
MN-Minneapolis
Site: MN-Minneapolis
Customer: Sono Bello
Customer #: 346
6545 France Ave South, Suite 302 Edina , MN 55435
Inspection #: 30216
Asset ID: 12972680
Model: Bovie - ARVV120
Notes: Electrical safety tested passed per NFPA 99 Specification [ 90 m⦠] Ground Resistance [ 3 µA ] Chassis Leakage (Norm/Cond) [ 12 µA ] Chassis Leakage (Open Gnd) Will not power on. Unit has power, but the on/off button will not turn the unit on.
Updating eBiotracks information
Cleaned device externals
Visual Inspection; Passed
Ground (earth) integrity (≤ 0.5 Ohms) = __0.12__ Ohms; passed
Earth leakage current test (≤ 300 µA) = __90__ µA; Passed
Electrical Safety Test; Passed
Status LEDs function - Passed
Test suction at all speeds (low / medium / high) - Passed
Filter life indicator - Passed
Footswitch - verified
🔍Technician Findings
Was not able to duplicate issue. Unit was able to power on after being plugged in with no issues. Once plugged in, unit's lights lit up not showing any sign or errors. Powered on unit and was able to activate each fan setting and filter is has suction pressure in all three inserts. Unit's filter still has life and does not need changed. Checked damages on unit and plastic cover around the front of the unit has some damages and cracks, but not enough to warrant necessary repairs. Preparing unit for Hotswap shelf for SONO BELLO.
⚠️Problem Description
MN-Minneapolis
Site: MN-Minneapolis
Customer: Sono Bello
Customer #: 346
6545 France Ave South, Suite 302 Edina , MN 55435
Inspection #: 30216
Asset ID: 12972680
Model: Bovie - ARVV120
Notes: Electrical safety tested passed per NFPA 99 Specification [ 90 m⦠] Ground Resistance [ 3 µA ] Chassis Leakage (Norm/Cond) [ 12 µA ] Chassis Leakage (Open Gnd) Will not power on. Unit has power, but the on/off button will not turn the unit on. https://www.ebiotrack.com/workorder.php?wo=71186
📄Description
Bovie - Smoke Shark Smoke II Evacuator w/ Power Cord and Bovie Smoke Filter (SONO BELLO)
Timeline
📞
Case Linked Nov 11, 12:26 PM
Case Number:00542932
Subject:eBiotrack - New Work Order Request - Inspection Repair
Status:Closed
🚨50.0 business days
📋
▶
3
Work Order Created Jan 27, 3:59 PM
Work Order Created → Parts Requested → Status: Awaiting Parts