eBiotrack - New Work Order Request - Inspection Repair
Closed
Oak Street Health - Cedar Crest
Site: Oak Street Health - Cedar Crest
Customer: Oak Street Health
Customer #: 256
2663 S. Lancaster Road Dallas, TX 75216
Inspection #: 31878
Asset ID: 13214916
Serial #: V2683875
Model: Midmark - 626-003
Notes: Electrical Safety Tested Passed Per NFPA 99 Specifications Ground Resistance: _291_____ (mOhms) Chassis Leakage (NC): _0_____ (uAmps) Chassis Leakage (SF): __128____ (uAmps) Failed Calibration Recommended Device Replacement Intermittent failure Follow-up required. Reviewed service with customer, quote needed Parts Needed- Emsar to follow up with quote for needed replacement part. Both foot and hand control have no function. Chair will need a wireless PCB, part# 002-10106-00 to complete repair.
2/26 KKelch scheduled per logged email from Amber. Email logged on WO-01349884
2/26 KKelch sent an email to Amber for scheduling. Email logged on WO-01349884
2/23 KKelch sent an email to Amber for rescheduling per CDempsey. Placed in R/S pending response. Email logged on WO-01349884
2/23 KKelch scheduled per logged email from Amber. Email logged on WO-01349884
2/23 KKelch sent Amber an email for scheduling. Email logged on WO-01349756
2/18/26 RBerry- Updated due date based on parts. Updated delay code.
2/18 KKelch sent Amber an email for scheduling. Email logged on WO-01349884
2/13 KKelch sent Amber an email for scheduling. Email logged on WO-01349884
🔧Work Performed
Installed new Wireless PCB Board. Unit function tested and passed. Unit returned to service. No travel do to multiple work orders at site.
🔍Technician Findings
3/3/2026 Zachary Stieben arrived on site. Found Midmark exam chair Asset# ID 13214916 with no wireless function.
⚠️Problem Description
Oak Street Health - Cedar Crest
Site: Oak Street Health - Cedar Crest
Customer: Oak Street Health
Customer #: 256
2663 S. Lancaster Road Dallas, TX 75216
Inspection #: 31878
Asset ID: 13214916
Serial #: V2683875
Model: Midmark - 626-003
Notes: Electrical Safety Tested Passed Per NFPA 99 Specifications Ground Resistance: _291_____ (mOhms) Chassis Leakage (NC): _0_____ (uAmps) Chassis Leakage (SF): __128____ (uAmps) Failed Calibration Recommended Device Replacement Intermittent failure Follow-up required. Reviewed service with customer, quote needed Parts Needed- Emsar to follow up with quote for needed replacement part. Both foot and hand control have no function. Chair will need a wireless PCB, part# 002-10106-00 to complete repair. https://www.ebiotrack.com/workorder.php?wo=71395
📧
Email:Re: Scheduling for WO-01349884
Completed Feb 23, 2026
From: kaitlyn.kelch@emsar.com
To: amber.valenzuela@oakstreethealth.com
Good morning Amber,
I just wanted to follow up on the below and see if 2/25 between 3 and 4 pm would work for your site to schedule these repairs.
Thank you, Kaitlyn KelchAccount Service Coordinator I-West District(937) 741-4389kaitlyn.kelch@emsar.comwww.emsar.comDisclaimer: The information in this e-mail is confidential and may be legally privileged; it is intended solely for the addressee. If you have received this c
📧
Email: Scheduling for repairs needed after inspection
Completed Feb 13, 2026
To: amber.valenzuela@oakstreethealth.com
CC:
BCC: kaitlyn.kelch@emsar.com
Attachment: --none--
Subject: Scheduling for repairs needed after inspection
Body:
Good morning Amber,
It looks like there are two devices that failed your inspection and they will need repairs performed on them. Our technician, Zac, can be there on 2/20 between 9 and 11 am to complete these. Please let me know if this date and time works for you.
Thank you,
Kaitlyn Kelch
Account Service Coordinator I-West District
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Feb 5, 2026
To: amber.valenzuela@oakstreethealth.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01349756Location: Oak Street Health - Cedar CrestAddress: 2663 S. Lancaster Road Dallas, TX 75216Device: Problem Description: Oak Street Health - Cedar Crest
Site: Oak Stre
📅
WO-01349756 - Virtual Tech - Field Repair - Dallas - Oak Street Health - Cedar Crest
Mar 3, 2026 10:00am – 12:00pm
Work Order: WO-01349756
Account: Oak Street Health
Customer Case Number:
Location: Oak Street Health - Cedar Crest
Address: 2663 S. Lancaster Road, Dallas, TX 75216
Manufacturer:
Model #:
Serial #:
Contact Name: Amber Valenzuela
Contact Phone:
Contact Email: amber.valenzuela@oakstreethealth.com