FSE Fady Khouzam - Upon initial inspection of the device, A pre‑existing fault was detected: the device failed the Active Exhalation Verification in the System Setup Test. This must be resolved before proceeding with FCO86600081 implementation.
03/16/2026 RArnold -Received email from customer today that they will have the PO to us by today
03/06/2026 RArnold - Fady M emailed customer for PO
03/06/2026 RArnold - Fady M emailed customer for PO
02/25/2026 RArnold - Per Fady K, Quote has been sent to the customer and we are waiting on the PO
02/17/2026 RArnold - PWO still pending entitlement
02/06/2026 RArnold - Repair for device that failed FCO81. PWO pending entitlement
🔧Work Performed
The FSE installed a new RP-Trilogy Evo 3 Way Solenoid Valve and RP-Trilogy Evo Proportional Valve (AECM), Performed a complete PVT as per the MFG's specifications. All tests and calibrations were found to be in spec or passed tests.Device placed back into full clinical service; ready for patient use.
🔍Technician Findings
Repair.
⚠️Problem Description
FSE Fady Khouzam - Upon initial inspection of the device, A pre‑existing fault was detected: the device failed the Active Exhalation Verification in the System Setup Test. This must be resolved before proceeding with FCO86600081 implementation.