The FSE installed a new RP-Trilogy Evo 3 Way Solenoid Valve and RP-Trilogy Evo Proportional Valve (AECM), Performed a complete PVT as per the MFG's specifications. All tests and calibrations were found to be in spec or passed tests.Device placed back into full clinical service; ready for patient use.
🔍Technician Findings
Repair.
⚠️Problem Description
Failed AECM test, needing 3 way solenoid and proportional valve
WO-01351522 - Repair - Trilogy EV300 Vent - Florence - Mcleod Regional Medical Center
Feb 12, 2026 8:00am – 5:00pm
Work Order: WO-01351522
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-14488129
Location: Mcleod Regional Medical Center
Address: 1145 Day St, Florence, SC 29506
Manufacturer: Philips Healthcare
Model #: Trilogy EV300 Vent
Serial #: H3034118942CC
Contact Name: Robert Fuller
Contact Phone: 843)
Timeline
📞
Case Linked Feb 13, 8:09 AM
Case Number:00568146
Subject:URGENT Philips WO Request
Status:Closed
📋
▶
4
Work Order Created Feb 13, 8:18 AM
Work Order Created → Technician Assigned → Dispatch Scheduled → Status: Scheduled
👷
Technician Assigned by Rebecca Arnold
Technician:Fady Khouzam
📅
Dispatch Scheduled by Rebecca Arnold
Scheduled For:2026-02-12T13:00:00.000+0000
🗓️
Status: Scheduled by Rebecca Arnold
From:Entered
To:Scheduled
Duration in Previous:3m
⏳1.0 business days
⏱️
2x Labor Added Feb 16, 9:17 AM
Type:Labor
Status:Open
Qty:1, 3
✔️
Status: Completed by Fady KhouzamFeb 16, 10:37 AM
From:Scheduled, Completed
To:Completed, Closed
Duration in Previous:3d 2h, 0m
⏳2.0 business days
⏱️
▶
2
Labor Added Feb 18, 1:40 PM
Labor Added → Status: 07e-Partial Project
📋
Status: 07e-Partial Project by Timothy Amburgey
From:Closed
To:07e-Partial Project
Duration in Previous:2d 3h
⏳2.0 business days
📄
Status: Ready to Bill by Timothy AmburgeyFeb 19, 4:31 PM
From:07e-Partial Project
To:Ready to Bill
Duration in Previous:1d 2h
💰
Status: Invoiced by Josh LongwayFeb 20, 4:01 AM
From:Ready to Bill
To:Invoiced
Duration in Previous:11h 29m
Bottlenecks
Closed
Duration: 2.0 business days (threshold: 1.0 business days)
Exceeded by 1.0 business days
Ready to Bill
Duration: 1.3 business days (threshold: 1.0 business days)