Zone 2
NWD Not turning on
ADDITIONAL NOTES
If part(s) are identified after speaking with Midmark Tech Service, the Service Company Tech must contact the POC for
authorization before ordering. Once approved, notify Midmark immediately. If more time is needed for the customer’s decision
to proceed, inform Midmark Dispatch at 937.688.9175 or dispatch@midmark.com. For assistance, troubleshooting, or parts
identification, contact Medical Tech Service at 844-856-1230. Thank you!
The following items have shipped from Midmark Corporation on: 23-FEB-2026
LINE NO: 1.1 QTY: 1 ITEM: 002-10958-00 SERIAL #: DESCRIPTION: KIT, BULB SOCKET AND HEAT SINK
2/17/26 AJinerson - Called and confirmed w/ Danielle to schedule for Monday, 2/23 with eta between 9am-11am.
POC onsite: Danielle
2/17/26 SReich - Spoke to Danielle, gave her the WO # and told her someone would be reaching out to schedule.
🔧Work Performed
Took switch cover and switch off to test voltage, tested voltage on wires going to switch, reading came back 27 voltage AC. Also took bulb socket off to test voltage where bulb connects, reading also came back 27 voltage AC. Tried a different bulb, bulb flicker but didn’t come on.
Contacted Midmark and spoke with Dale of my findings, determined issue is with bulb socket and will need to be replaced. Cost for replacement bulb socket kit is $106, received approval from Iva to order replacement. Midmark is going to ship replacement directly to customer.
🔍Technician Findings
Tried turning light on from switch on light and wouldn’t turn on, noticed one side of switch cover cracked. Light only controlled by switch on light.
⚠️Problem Description
Zone 2
NWD Not turning on
ADDITIONAL NOTES
If part(s) are identified after speaking with Midmark Tech Service, the Service Company Tech must contact the POC for
authorization before ordering. Once approved, notify Midmark immediately. If more time is needed for the customer’s decision
to proceed, inform Midmark Dispatch at 937.688.9175 or dispatch@midmark.com. For assistance, troubleshooting, or parts
identification, contact Medical Tech Service at 844-856-1230. Thank you!
Work Order: WO-01352381
Account: MIDMARK
Customer Case Number: SR 1539330
Location: Forever Young
Address: 1485 Pine Ridge Rd, Unit 1, Naples, FL 34109-2114
Manufacturer: Midmark
Model #: 300 (Series) Lights
Serial #: V1165704
Contact Name: Danielle Cullen
Contact Phone: 16367510031
Contact Email: d