Note
02/17/2026 12:39:28 WIGEORGE
gratis replacement following pm call last week.
____________________
Approval Note
02/17/2026 12:39:28 WIGEORGE
wesley McQeuen
____________________
Note
02/17/2026 12:28:50 WF-BATCH
Agent: WIGEORGE Timestamp: 2026-02-17 12:28:42
sales Ron Neikirk ron.neikirk@agilent.com mobile: 360-281-3249
gratis field repair; fils out from service last week. 6007704686
3-4 Mburris tech is needed on a big job in Sac / The Sac and Hillsboro, OR POC got mixed up so Hillsboro, OR is getting pushed out a week to accommodate with Sac sites urgency
3/2/26 RBerry- Updated due date based on parts for WO-01352637 and WO-01352632. Site wants these scheduled for the same trip. States they were to be sent next day and wanted these on the calendar. Chattered LCoonrod on WO-01352637 and WO-01352632 to confirm parts.
2-26 Mburris called spoke with Ian MacLeod. Advised that an email came over with wrong wo's for his site He set appts for 3-10 at 7am for Hillsboro, Or 3 wo's from email show Sac, CA / also informed he that site has 5 WO's POC said all parts were sent over next day and that he also has extra parts on site for tech to use.
2/26/26 lcoonrod Part still showing in open status. But per email from Ian - I have the parts . Please schedule ASAP
2/23/26 lcoonrod parts are still showing in open status.
2/20/26 lcoonrod Parts have not shipped yet
2/17/26 lcoonrod Parts have not shipped yet
VSFLDHIS HOT ION SOURCE KIT
VSFLDOV VS SERIES, VALVE BLOCK O-RING KIT
🔧Work Performed
Completed Service Tasks:
Inspection and Cleaning
Performed a comprehensive inspection and cleaning of:
Spectrometer
Valve block
Manifold
Component Replacement:
Replaced ion source and filaments
Replaced o-rings in diaphram pump.
Installed new O-rings and seals where required
Replaced tip seal in vacuum pump.
System Verification:
Verified integrity of the vacuum system
Performed helium sensitivity check
Completed full system calibration
Calibrated Leak Maintenance:
Replaced the calibrated leak
Supplied ISO traceable certificate for the new calibrated leak
🔍Technician Findings
As found: Unit in workshop.
⚠️Problem Description
Note
02/17/2026 12:39:28 WIGEORGE
gratis replacement following pm call last week.
____________________
Approval Note
02/17/2026 12:39:28 WIGEORGE
wesley McQeuen
____________________
Note
02/17/2026 12:28:50 WF-BATCH
Agent: WIGEORGE Timestamp: 2026-02-17 12:28:42
sales Ron Neikirk ron.neikirk@agilent.com mobile: 360-281-3249
gratis field repair; fils out from service last week. 6007704686
📄Description
Helium Leak Detector, MD15, Dry Scroll IPD-15
📞Call Description
VPD Field Repair Trade
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Feb 25, 2026
To: ian.mcleod@thermofisher.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01352449Location: FEI CoAddress: 5350 NE Dawson Creek Dr Hillsboro, OR 97124Device: G8611C MY2119HD02Problem Description: Note
02/17/2026 12:39:28 WIGEORGE
gratis replacement follo
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Feb 18, 2026
To: ian.mcleod@thermofisher.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01352449Location: FEI CoAddress: 5350 NE Dawson Creek Dr Hillsboro, OR 97124Device: G8611C MY2119HD02Problem Description: Note
02/17/2026 12:39:28 WIGEORGE
gratis replacement follo
Work Order: WO-01352449
Account: AGILENT TECHNOLOGIES (VPD-FIELD)
Customer Case Number: 6008466454
Location: FEI Co
Address: 5350 NE Dawson Creek Dr, Hillsboro, OR 97124
Manufacturer:
Model #: G8611C
Serial #: MY2119HD02
Contact Name: Ian MacLeod
Contact Phone: 503-619-7642
Contact Email:
PDF
HLD PM Checklist A.00 (1)
Feb 17, 2026
PDF
HLD Enhanced PM Checklist A.01 (1)
Feb 17, 2026
PDF
HLD PM Checklist A.00 - WO-01352449
Mar 22, 2026
Timeline
📞
Case Linked Feb 17, 12:57 PM
Case Number:00569249
Subject:6008466454 FW: Repair request
Status:Closed
📋
▶
3
Work Order Created Feb 17, 4:29 PM
Work Order Created → Status: Awaiting Parts → 2x Parts Added
📦
Status: Awaiting Parts by Leona Coonrod
From:Entered
To:Awaiting Parts
Duration in Previous:0m
🔩
2x Parts Added
Type:Parts
Status:Open
Qty:1
⚠️7.9 business days
📅
Status: Ready for Scheduling by Leona CoonrodFeb 26, 2:42 PM