Re: West Palm Beach: Equipment Issue – Pressure Tube and Hub Sensor
Closed
Suntech Tango M2
It appears the pressure tube is leaking, and the hub cable/sensor is not functioning properly.
This unit will need to be sent to the depot for services to be completed.
2hr depot labor
1x - Suntech Original NIBP Hose - 91-0127-01
3x shipping @30
Assisted with evaluation of Air leak and serial # of Monitor. Air Leak on back port of unit either connection port or hose on inside of unit. Sending Case to Depot to Quote.
🔍Technician Findings
Air Leak on Air port on back of Monitor. Recommended Depot Repair.
⚠️Problem Description
-- Please reply above this line --
Hello, The clinic below has a Suntech Tango M2 that is giving them
problems, is this something that you are familiar with and can help
them with? We will need a quote before any repairs are made.
_"It appears the pressure tube is leaking, and the hub cable/sensor is
not functioning properly. Due to this issue, we may need a replacement
part or a technician to assess and repair the equipment."_
SITE NAME: City of West Palm Beach - SITE ID: 22607 - SMX ID:
a0sVU0000013q2lYAA
CUSTOMER NAME: CareATC
SITE ADDRESS: 413 Clematis St, #230, Palm Beach FL 33401
[image:
https://d33v4339jhl8k0.cloudfront.net/inline/69283/625e0b08c1cfd7f94389ab06aad1580f419e4e0d/09cecb5e15fa71ef9b9e4b7f758acd8c970728d1/image.png]
--
Your ticket number is: #359172
Jennifer Rice
purchasingsupport@careatc.com
How would you rate my reply?
Great [1] Okay [2] Not Good [3]
Links:
------
[1]
https://secure.helpscout.net/satisfaction/765213941/record/9706453062/1/
[2]
https://secure.helpscout.net/satisfaction/765213941/record/9706453062/2/
[3]
https://secure.helpscout.net/satisfaction/765213941/record/9706453062/3/
> On Tue, Dec 30, 2025 at 4:09 PM CST, Mimi Johnson <mimijohnson@careatc.com> wrote:
>
> ** CAUTION ** This email originated from outside of CareATC. Do not
> click links, open attachments or follow guidance unless you recognize
> the sender and know the content is safe.
> Good afternoon, It appears the pressure tube is leaking, and the hub
> cable/sensor is not functioning properly. Due to this issue, we may
> need a replacement part or a technician to assess and repair the
> equipment. Please advise on next steps. Thank you, MIMI JOHNSON
> Center Operations Manager
>
> CAREATC, INC.
> 413 Clematis St, 230
> West Palm Beach, FL 33401
> O: 561.822.2000
> www.CareATC.com [1]
>
> [image:
> https://d33v4339jhl8k0.cloudfront.net/inline/69283/9225456657a2a63a7a8da98912ad40003fd61879/25a6454036d6287a2b73f862d9625fc6c868836b/Outlook-CareATC.jpg]
>
>
>
> Links:
> ------
> [1] https://www.careatc.com/
>
>
>
> https://www.ebiotrack.com/workorder.php?wo=70630
SNOTE
VT Notes
Dec 31, 2025
Case #00555304 – Re: West Palm Beach: Equipment Issue – Pressure Tube and Hub Sensor
Status: Closed | Priority: Standard
Suntech Tango M2
It appears the pressure tube is leaking, and the hub cable/sensor is not functioning properly.
This unit will need to be sent to the depot for services to be completed.
2hr depot labor
1x - Suntech Original NIBP Hose - 91-0127-01
3x shipping @30
💬Case Comment
Internal Jan 5, 2026 3:13pm
reached out to orders to add part to SM
Timeline
📞
Case Linked Dec 31, 10:33 AM
Case Number:00555304
Subject:Re: West Palm Beach: Equipment Issue – Pressure Tube and Hub Sensor