Cleaned device externals
Cleared stored ePHI
Visual Inspection; Passed
AM12 Condition; Good (S/N:______123150002494 ______)
Battery Date Code = ___2025____
Battery Test - No Load - (> 12.5VDC) = ___12.88____VDC
Battery Test - 10Ω25W - (>11.7VDC) after 10 sec = ____12.43___VDC
Current Test - Power OFF - (< 250uA) = ___113.64____uA
Current Test - Power ON - (< 1.1A) = ___0.53____A
Current Test - Charging voltage (13.0 - 14.7 VDC) = ___13.64____VDC
Touchscreen Calibrated
Date/Time; Updated
Auto Test; Passed
Writer Test; Passed
Printhead in good condition, cleaned, Que verified
Performance Tested EKG Leads @ 30, 60, 80, 120 BPM; Passed
Lead-off Warning; Verified
Noise Test; Passed
Electrical Safety Test; Passed
Enclosure Leakage Current measured >300µA; Passed
Patient Leakage Current measured >10µA; Passed
PM Completed
🔍Technician Findings
Was not able to duplicate issue. AM12 was able to connect to device and record readings. Lead wires provided with unit are in good working condition. Unit did not come with clips and will need to be added to lead wires. Unit will also need:
1 x 3600-016
1 x 715006
1 x RPC260
1 x 9100-026-01
⚠️Problem Description
Work Order #98306
EKG Machine- It's not staying charged and not registering the leads
Burdick - ELI 280
Asset ID#: 13095099
S/N: 122110000076 https://www.ebiotrack.com/workorder.php?wo=72257
📄Description
ECG, BURDICK 280 INTERP 200 ECG STORAGE AM12/USB/LAN
Oak Street Health’s Consigned Inventory
Timeline
📞
Case Linked Oct 2, 11:37 AM
Case Number:00532447
Subject:Oak Street Health - 48 Hour - 98306: New Work Order Assigned
Status:Closed
🚨113.0 business days
📋
▶
2
Work Order Created Mar 9, 9:48 AM
Work Order Created → Status: Assigned
👷
Status: Assigned by Samuel Klein
From:Tech On Site
To:Assigned
Duration in Previous:0m
🔧
Status: Tech On Site by Samuel KleinMar 10, 8:09 AM
From:Assigned
To:Tech On Site
Duration in Previous:22h 21m
🔩
6x Parts Added Mar 10, 8:18 AM
Type:Parts, Labor
Status:Open
Qty:1
✔️
Status: Completed by Samuel KleinMar 10, 8:52 AM
From:Tech On Site, Completed
To:Completed, Closed
Duration in Previous:43m, 0m
⏳2.0 business days
📄
Status: Ready to Bill by Diane PattonMar 11, 11:42 AM
From:Closed
To:Ready to Bill
Duration in Previous:1d 2h
🚨16.0 business days
📨
Status: Invoice Pending by Diane PattonApr 1, 5:31 PM
From:Ready to Bill
To:Invoice Pending
Duration in Previous:21d 5h
🧾
▶
2
Invoice Created Apr 1, 6:10 PM
Invoice Created → Status: Invoiced
💰
Status: Invoiced by Diane Patton
From:Invoice Pending
To:Invoiced
Duration in Previous:39m
Bottlenecks
Assigned
Duration: 2.0 business days (threshold: 4.0 business hours)
Exceeded by 1.5 business days
Closed
Duration: 2.0 business days (threshold: 1.0 business days)
Exceeded by 1.0 business days
Ready to Bill
Duration: 16.0 business days (threshold: 1.0 business days)