Oak Street Health - 48 Hour - 124603: New Work Order Assigned
Closed
Work Order #124603
Work Order Description:
BP Machine giving error stating CSM battery is not installed 1c100e. If machine is not plugged in, it doesn't allow MA to use it. Once plugged it works but same message appears.
Cleaned Device Externals
Visual Inspection; Passed
Battery Replaced
Initialize NIBP Board Completed
Calibration Completed
PM Completed per mfg specs
*See attached service record for detailed testing results*
🔍Technician Findings
Was able to duplicate issue. Unit was working through first calibration and then showed an error #00001 code indicating that the PCB board is bad and will need replaced. Unit came in with all accessories except:
1 x Power Supply (7000-PS)
and
unit will need:
1 x Refurbished PCB Board (105925-R)
1 x Batt22
⚠️Problem Description
Work Order #124603
BP Machine giving error stating CSM battery is not installed 1c100e. If machine is not plugged in, it doesn't allow MA to use it. Once plugged it works but same message appears.
Welch Allyn - 71XX - Connex Spot Monitor
Asset ID#: 13353433
S/N: 21100053991321
Department: Primary Care
Physical Location: 6 https://www.ebiotrack.com/workorder.php?wo=72477
📄Description
Welch Allyn - Connex Spot Monitor 71XX
Oak Street Health Consignment Equipment