Oak Street Health - 48 Hour - 124603: New Work Order Assigned
Closed
Work Order #124603
Work Order Description:
BP Machine giving error stating CSM battery is not installed 1c100e. If machine is not plugged in, it doesn't allow MA to use it. Once plugged it works but same message appears.
Cleaned Device Externals
Visual Inspection; Passed
Battery Replaced
Calibration Completed
PM Completed per mfg specs
*See attached service record for detailed testing results*
🔍Technician Findings
Was able to duplicate issue. Battery in unit has gone bad and will need a new battery to fix issue. After battery is replaced, performing PM on unit to ensure unit is running correctly with no issues and placing onto Hotswap shelf.
Unit will need:
1 x Batt22
1 x Reuse-11
⚠️Problem Description
Work Order #124603
BP Machine giving error stating CSM battery is not installed 1c100e. If machine is not plugged in, it doesn't allow MA to use it. Once plugged it works but same message appears.
Welch Allyn - 71XX - Connex Spot Monitor
Asset ID#: 13353409
S/N: 21100008062325
Department: Primary Care
Physical Location: 6 https://www.ebiotrack.com/workorder.php?wo=72824
📄Description
Welch Allyn - Connex Spot Monitor 71XX
Oak Street Health Consignment Equipment